The development and implementation of phased array technology for small animal MRI systems provides superior signal-to-noise ratio compared to standard quadrature coils.
Answers are grounded in this paper and cite their sources.
Array coils may enhance small-animal MRI; extends coil designs but leaves human translation open.
Although phased array technology has been standard on clinical magnetic resonance imaging (MRI) systems for many years, it is just now becoming available on small animal MRI systems. Bruker Instruments Ltd. has recently provided multiple rf channels on a 4.7T/33cm Avance, but further development was necessary to complete the phased array implementation. This work discusses the development of this other hardware, specifically the rf array coil, low impedance preamplifiers, excitation coil, and image combination algorithm. Comparison of the array coil to a quadrature coil indicates superior signal-to-noise ratio from the array. In vivo images of a cat spine acquired simultaneously from the individual channels of the array and a sum of squares combination are shown.
No takes yet. Share an insight, caveat, or question.
Beck et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: