Key result
A non-intrusive pneumatic sensor accurately estimated overnight cardiorespiratory metrics compared to polysomnography, demonstrating a heart rate bias of -0.8 bpm (95% LOA -4.3 to 2.7).
Why the study?
Does a non-intrusive pneumatic sensor accurately measure cardiorespiratory metrics during sleep compared to polysomnography in sleep laboratory participants?
Observational (n=42)
Does a non-intrusive pneumatic sensor accurately measure cardiorespiratory metrics during sleep compared to polysomnography in sleep laboratory participants?
A non-intrusive pneumatic sensor can accurately estimate overnight cardiorespiratory metrics including heart rate and respiratory rate compared to standard polysomnography.
Answers are grounded in this paper and cite their sources.
May facilitate accessible overnight monitoring in research; leaves open home validation and clinical outcome impact.
We show that a non-intrusive pneumatic sensor can be used to measure respiratory rate, heart rate, and their variability during sleep. The pneumatic sensor was included in a polysomnography (PSG) study involving 42 participants in a sleep laboratory. The agreement between the pneumatic sensor and the PSG for respiratory rate, heart rate, and their variability was quantified by Bland-Altman analysis. The respiratory rate has a mean value of 15.4 breaths per minute for a bias of −0.06 and 95% limits of agreement (LOA) of [−0.6; 0.5] breath per minute. The respiratory rate variability root mean square of successive differences (RMSSD) has a mean value of 459.51 ms, a bias of 9.2 and a 95% LOA of [−43.5; 61.9] ms. The heart rate has a mean value of 60.6 beats per minute for a bias of −0.8 and a 95% LOA of [−4.3; 2.7] beat per minute. The heart rate variability RMSSD has a mean value of 44.1 ms, a bias of 14.7 and a 95% LOA of [−19; 48.4] ms. These results show that a non-intrusive pneumatic sensor can accurately estimate cardiorespiratory metrics overnight.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2019) conducted an observational in Sleep (n=42). Non-intrusive pneumatic sensor vs. Polysomnography (PSG) was evaluated on Agreement between pneumatic sensor and PSG for respiratory rate, heart rate, and their variability (Bias -0.8 bpm (heart rate), 95% CI -4.3 to 2.7 (95% LOA)). A non-intrusive pneumatic sensor accurately estimated overnight cardiorespiratory metrics compared to polysomnography, demonstrating a heart rate bias of -0.8 bpm (95% LOA -4.3 to 2.7).