Key result
Home blood pressure monitoring showed no significant difference in nocturnal BP compared to ambulatory monitoring, suggesting it may be a reliable alternative for assessing nocturnal BP.
Why the study?
Nocturnal BP strongly predicts hypertensive target organ damage, but the agreement between ambulatory BP monitoring and home BP monitoring for nocturnal BP assessment in a general population required evaluation.
Does home blood pressure measurement reliably assess nocturnal blood pressure compared to ambulatory blood pressure monitoring in a general population?
Observational (n=55)
Does home blood pressure measurement reliably assess nocturnal blood pressure compared to ambulatory blood pressure monitoring in a general population?
Home blood pressure monitoring may be a reliable alternative to ambulatory blood pressure monitoring for assessing nocturnal blood pressure.
Answers are grounded in this paper and cite their sources.
May support home BP as nocturnal assessment option; leaves open prospective validation before practice change.
OBJECTIVES: The nocturnal blood pressure (BP) is a strong predictor of hypertensive target organ damage including that in cardiovascular diseases. The use of ambulatory BP (ABP) monitoring has enabled the evaluation of nocturnal BP and detection of non-dippers. This study compared nocturnal BP values, nocturnal decline in BP, and the prevalence of non-dippers based on ABP and home BP (HBP) measurements in a general population. METHODS: Data on HBP measured with HEM 747-IC-N (Omron Healthcare Co., Ltd.) and 24-hour ABP measured with ABPM-630 (Nippon Colin) were obtained from fifty-five participants aged ≥ 20 years (mean age: 65.1 years, 78.2% women). To exclude a systematic difference between the two methods, we conducted a validation study for HBP and ABP in another population that consisted of hypertensive outpatients (mean age: 65.4 years, 53.4% women). RESULTS: After adjusting for the systematic difference in BP between the two methods calculated in the validation study (3.9 mmHg for systolic and 3.0 mmHg for diastolic), morning and daytime (average of morning and evening) HBP were significantly lower than morning (average of 2 h after waking) and daytime (average of being awake) ABP, respectively. No significant difference was found in nocturnal BP between HBP and ABP monitoring regardless of the quality of sleep during nocturnal HBP measurement. Agreement between HBP and ABP in the detection of non-dippers was low mainly due to the difference in daytime BP values. Conclusion: HBP monitoring may be a reliable alternative to ABP for the assessment of nocturnal BP.
No takes yet. Share an insight, caveat, or question.
Hosohata et al. (2020) conducted an observational in General population (n=55). Home blood pressure (HBP) measurement vs. Ambulatory blood pressure (ABP) measurement was evaluated on Nocturnal BP values, nocturnal decline in BP, and prevalence of non-dippers. Home blood pressure monitoring showed no significant difference in nocturnal BP compared to ambulatory monitoring, suggesting it may be a reliable alternative for assessing nocturnal BP.