Key result
A data-fusion algorithm combining ultra-wideband and IMU signals achieved an error rate for respiration rate estimation of less than 0.2 rpm compared to a reference system.
A novel data-fusion algorithm combining UWB and IMU sensors enables highly accurate respiration rate monitoring with an error rate <0.2 rpm, suitable for wearable devices.
This paper presents a data-fusion algorithm for respiration rate (RR) extraction by employing the ultra-wideband (UWB) transversal propagation measurement method. In the experimental scenario the UWB transmitter and receiver units were placed on the front and back sides of the thoracic wall. The measurement principle is based on the fact that periodic movements of lung affect the communication channel properties that can be measured. We measured the energy attenuation variations on the receiver side caused by tissue movements from which we extracted the information about the respiration rate. For testing purpose, a custom developed UWB platform suitable for on-body placement was used, with an integrated inertial motion unit (IMU) sensor. UWB and IMU signals were combined by means of a data fusion algorithm. Data fusion algorithms based on Extended Kalman filtering (EKF) and Naïve Bayes inference show better estimation performance than an estimation from individual signal sources. The obtained error rate of RR estimation by means of the proposed data fusion method is lower than 0.2 respiration per minute (rpm) in comparison to the reference respiration system. Our results show that the proposed method of UWB and IMU sensor fusion is a promising candidate for reliable RR monitoring by compact wearable units under relatively non-stationary body movement conditions.
No takes yet. Share an insight, caveat, or question.
Čuljak et al. (2020) studied Respiration rate monitoring. Data-fusion algorithm combining UWB and IMU signals vs. Individual signal sources and reference respiration system was evaluated on Error rate of respiration rate (RR) estimation. A data-fusion algorithm combining ultra-wideband and IMU signals achieved an error rate for respiration rate estimation of less than 0.2 rpm compared to a reference system.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: