Answers are grounded in this paper and cite their sources.
An autocorrelation-based algorithm using IR-UWB radar provides a highly accurate, non-contact method for measuring respiration and heartbeat rates.
Respiration rate (RR) and heartbeat rate (HR) are important physiological parameters for a person. Impulse radio ultra-wideband (IR-UWB) is a promising technology for non-contact sensing and monitoring. This brief presents a new method based on autocorrelation to measure the RR and HR using IR-UWB radar. The correlation coefficient waveform contains the vital sign signals, overcoming the effect of noise and clutter. Applying fast Fourier transform, the respiration frequency can be acquired easily. A clever method also based on autocorrelation is proposed to locate the subject. The receive signal matrix is divided into a set of bins in the direction of fast time. By removing one block from the matrix each time and re-applying the autocorrelation, the removed block resulting the smallest correlations is corresponding to the location of a subject. Moreover, variational mode decomposition algorithm is adopted to successfully separate the respiration and heartbeat signals. Experiments are carried out using a PulsOn410 UWB radar. The results show that the proposed low-complexity algorithm has high accuracy.
No takes yet. Share an insight, caveat, or question.
Shen et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: