Key result
Combining complex signal demodulation and arctangent demodulation techniques with a state-space method improved the signal-to-noise ratio for heart rate detection by over 10 dB.
Answers are grounded in this paper and cite their sources.
Advanced signal demodulation techniques combined with state-space methods significantly improve the signal-to-noise ratio for noncontact heart rate detection using UWB radar.
Ultra-wideband (UWB) pulse Doppler radars can be used for noncontact vital signs monitoring of more than one subject. However, their detected signals typically have low signal-to-noise ratio (SNR) causing significant heart rate (HR) detection errors, as the spurious harmonics of respiration signals and mixed products of respiration and heartbeat signals (that can be relatively higher than heartbeat signals) corrupt conventional fast Fourier transform spectrograms. In this paper, we extend the complex signal demodulation (CSD) and arctangent demodulation (AD) techniques previously used for accurately detecting the phase variations of reflected signals of continuous wave radars to UWB pulse radars as well. These detection techniques reduce the impact of the interfering harmonic signals, thus improving the SNR of the detected vital sign signals. To further enhance the accuracy of the HR estimation, a recently developed state-space method has been successfully combined with CSD and AD techniques and over 10 dB improvements in SNR is demonstrated. The implementation of these various detection techniques has been experimentally investigated and full error and SNR analysis of the HR detection are presented.
No takes yet. Share an insight, caveat, or question.
Ren et al. (2016) studied Heart rate detection. Complex signal demodulation (CSD) and arctangent demodulation (AD) combined with state-space method vs. Conventional fast Fourier transform spectrograms was evaluated on Signal-to-noise ratio (SNR) (>10 dB improvement). Combining complex signal demodulation and arctangent demodulation techniques with a state-space method improved the signal-to-noise ratio for heart rate detection by over 10 dB.
Synapse has enriched 10 related papers. Consider them for comparative context: