Key result
Tissue Doppler imaging-derived strain and strain rate measurements are feasible in very low birth weight infants, with intra- and inter-operator variability ranging from 15.1% to 65.2%.
Why the study?
What are the optimal analysis parameters for tissue Doppler-derived strain and strain rate measurements in very low birth weight infants?
Observational
What are the optimal analysis parameters for tissue Doppler-derived strain and strain rate measurements in very low birth weight infants?
Tissue Doppler imaging-derived strain and strain rate measurements are feasible in very low birth weight infants, with optimal reproducibility achieved using specific region of interest and strain length parameters.
Answers are grounded in this paper and cite their sources.
High variability warrants caution interpreting tissue Doppler strain in very low birth weight infants; leaves optimal parameters open.
AIMS: To assess the effects of region of interest (ROI) size and strain length (SL) on the quality of analysis and to determine the feasibility and reproducibility of tissue Doppler-derived cardiac strain (S) and strain rate (SR) measurements for preterm infants <1500 g birth weight. METHODS AND RESULTS: The beat-to-beat variation (BBV) for different combinations of ROI width (RW), ROI length (RL), and SL was determined from 60 good quality images of left and right free walls and the interventricular septum. We examined the impact of RW, RL, and SL on BBV and estimated inter- and intra-operator variability. BBV was smallest for RW = 2 mm. Raising SL (even at the cost of RL) led to a decrease in BBV. Several combinations provided a low BBV without significant difference. Coefficients of variation for intra- and inter-operator variability ranged from 15.1% to 65.2%. Both were lower for systolic than for diastolic values. CONCLUSION: Tissue Doppler imaging-derived S and SR measurements are feasible in very low birth weight infants. Reproducibility is comparable to older infants for systolic S and SR. For optimal analysis quality, SL and RW should be high while RL should be small (though inside segment size). Best results were obtained with RW = 2 mm, RL = 1 mm, and SL = 6 mm.
No takes yet. Share an insight, caveat, or question.
Helfer et al. (2013) conducted an observational in Preterm infants <1500 g birth weight. Tissue Doppler-derived cardiac strain and strain rate measurements was evaluated on Beat-to-beat variation (BBV) and inter- and intra-operator variability. Tissue Doppler imaging-derived strain and strain rate measurements are feasible in very low birth weight infants, with intra- and inter-operator variability ranging from 15.1% to 65.2%.