Key result
A model-based estimation and actuation compensation framework successfully accounted for friction-induced extensions and cross-coupling effects in a two-segment single port access surgery robot.
The proposed framework enables online estimation of modeling uncertainties and friction in multisegment continuum robots, potentially improving their precision in surgical applications.
May enhance precision of experimental continuum robots; leaves open clinical translation beyond animal models.
The prevalence of wire-actuated mechanisms and continuum robots (CRs) for surgical applications stems from several advantages due to the remote location of actuators from the end-effectors (e.g., improved down scalability and sterilization). These advantages, however, come at the expense of inherent uncertainties due to backlash effects, compliance, and friction in the actuation lines, which in turn, limit their precision. In addition, multisegment CRs suffer from actuation coupling, diminishing the accuracy of their kinematic model. This paper aims to address these two gaps by presenting a model-based estimation and actuation compensation framework enabling the online estimation of modeling uncertainties and friction, despite possible temporal changes in these parameters and cross coupling between segments. A capstan friction model accounting for friction-induced actuation line extensions in both fixed and variable geometry conduits is presented. A modified statics model for multisegment robots is presented to account for cross-coupling effects between subsequent CR segments. A sequential estimation approach using the robot inverse kinematics, statics, and measurements of actuated joint positions and forces is then presented. These approaches are then validated experimentally on a two-segment single port access surgery CR.
No takes yet. Share an insight, caveat, or question.
Roy et al. (2016) studied Surgical applications (continuum robots). Model-based estimation and actuation compensation framework was evaluated. A model-based estimation and actuation compensation framework successfully accounted for friction-induced extensions and cross-coupling effects in a two-segment single port access surgery robot.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: