Key result
Simulated bipolar radiofrequency ablation at 25 W optimizes lesion performance in deep myocardium.
Why the study?
Myocardial baseline impedance affects ablation effectiveness, prompting the need to identify suitable bipolar radiofrequency ablation protocols and coping strategies under different baseline impedances.
Does low-power and long-duration bipolar radiofrequency ablation offer optimal performance under various baseline impedances in a myocardial computer model?
Does low-power and long-duration bipolar radiofrequency ablation offer optimal performance under various baseline impedances in a myocardial computer model?
Bipolar radiofrequency ablation at 25 W for 120 seconds provides optimal effectiveness and safety for 10 mm depth myocardium in computer simulations.
Simulation data favor LPLD bipolar RFA; hypothesis-generating and should not yet change practice.
• A validated bipolar ablation computer model and ablation criteria are presented. • The maximum TID decreased by 21–32 Ω in LPLD. • Bipolar 25 W for 120 s offered optimal performance for 10 mm depth myocardium. • The greater the power was, the less was the decrease in the maximum TID. Compared to traditional unipolar radiofrequency ablation (RFA), bipolar RFA offers advantages such as more precise heat transfer and higher ablation efficiency. Clinically, myocardial baseline impedance (BI) is one of the important factors affecting the effectiveness of ablation. We aim at finding suitable ablation protocols and coping strategies by analyzing the ablation effects and myocardial impedance changes of bipolar RFA under different BIs. In this research, a three-dimensional local myocardial computer model was constructed for bipolar RFA simulation, and in vitro experimental data were used to validate accuracy. Four fixed low-power levels (20 W, 25 W, 30 W, and 35 W) and six myocardial BIs (91.02 Ω, 99.83 Ω, 111.03 Ω, 119.77 Ω, 130.03 Ω, and 135.45 Ω) were set as initial conditions, with an ablation duration of 120-s. In the context of low-power and long-duration (LPLD) ablation, the maximum TID (TID M ) decreased by 21–32 Ω, depending on the BI. In cases where steam pop did not occur, TID M increased with the increase in power. For the same power, there was no significant difference in TID M for the range of BIs. In cases where steam pop occurred, for every 1 Ω increase in BI, TID M increased by 0.34–0.41 Ω. The simulation results also showed that using a higher power resulted in a smaller decrease in TID M . This study provided appropriate ablation times and impedance decrease ranges for bipolar LPLD RFA. The combination of 25 W for 120-s offered optimal performance when considering effectiveness and safety simultaneously.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2024) studied Radiofrequency ablation. Bipolar radiofrequency ablation was evaluated on Maximum total impedance drop (TID M). In a computer simulation of bipolar radiofrequency ablation, 25 W for 120 seconds offered optimal performance for 10 mm depth myocardium, with the maximum total impedance drop decreasing by 21–32 Ω.