Key result
Programming a monitoring zone in ICD patients detected arrhythmias in 62.9% of patients, prompting outpatient medication changes in 41.7% of those with reported events.
Why the study?
It was not known for certain whether programming an ICD monitoring zone affects arrhythmia detection.
Does programming a monitoring zone in ICDs improve the detection of arrhythmias and guide clinical management in patients with primary prevention ICDs?
Observational (n=221)
Does programming a monitoring zone in ICDs improve the detection of arrhythmias and guide clinical management in patients with primary prevention ICDs?
Programming a monitoring zone in ICDs frequently detects ventricular and supraventricular arrhythmias, which can usefully guide changes in outpatient medical therapy.
Answers are grounded in this paper and cite their sources.
May support arrhythmia surveillance and medication titration in primary-prevention ICDs; leaves open effects on shocks or survival.
INTRODUCTION: Implantable cardioverter-defibrillator (ICD) monitoring zones (MZ) provide passive features that do not interfere with the functioning of active treatment zones. However, it is not known for certain whether programming an MZ affects arrhythmia detection by the ICD. The aim of the present study is to assess the clinical relevance of MZ in a population of patients with ICDs. METHODS: In this retrospective analysis of patients with ICDs, with or without cardiac resynchronization therapy, for primary prevention under remote monitoring, the MZ was analyzed and recorded arrhythmias were assessed in detail. RESULTS: A total of 221 patients were studied (77% men; age 64±12 years). Mean ejection fraction was 30±12%. The mean follow-up was 63±35 months. One hundred and seventy-four MZ events were documented in 139 patients (62.9%): 74 of non-sustained ventricular tachycardia (NSVT), 42 of supraventricular tachycardia, 44 of atrial fibrillation/atrial flutter, and five cases of noise. Among the 137 patients who presented with arrhythmias in the MZ (excluding two cases with noise detection only), 22 (16.1%) received appropriate shocks and/or antitachycardia pacing (ATP), while of the other 84 patients, 15.5% received appropriate ICD treatment (p=NS). In patients who presented with NSVT in the MZ, 15 (20.5%) received appropriate shocks and/or ATP. In accordance with the MZ findings, physicians decided to change outpatient medication in 41.7% of all patients in whom arrhythmic events were reported. CONCLUSION: Ventricular and supraventricular arrhythmias are common findings in the MZ of ICD patients. Programming an MZ is valuable in the diagnosis of arrhythmias and may be a useful tool in clinical practice.
No takes yet. Share an insight, caveat, or question.
Rosa et al. (2019) conducted an observational in Implantable cardioverter-defibrillator (ICD) for primary prevention (n=221). Monitoring zones (MZ) programming vs. Patients without arrhythmias in the MZ was evaluated on Appropriate shocks and/or antitachycardia pacing (ATP) (p=NS). Programming a monitoring zone in ICD patients detected arrhythmias in 62.9% of patients, prompting outpatient medication changes in 41.7% of those with reported events.
Synapse has enriched 10 related papers. Consider them for comparative context: