A ventricular unipolar myocardial electrode can cause inadvertent atrial pacing, potentially leading to atrial arrhythmias.
Raises possibility of inadvertent atrial capture with unipolar ventricular leads; leaves open its contribution to device-related atrial arrhythmias.
A patient with sinoatrial node disease is reported who demonstrated inadvertent atrial pacing from a ventricular unipolar myocardial electrode and a VVI Intermedics pacemaker. Atrial refractory and vulnerable periods are addressed in this context as possibly providing the environment for an atrial arrhythmia, secondary to competition between the non-sensed impulse and the unipolar cathodal stimulus falling in the atrial vulnerable period.