Key result
Inadvertent atrial electrode placement causes epicardial ventricular pacing failure in a revised lead system.
Why the study?
Inadvertent atrial pacing due to epicardial electrode placement is an unusual finding that requires systematic analysis to avoid misplacement.
Case Report (n=1)
This case highlights the importance of systematic analysis of pacing system function at the time of implant to avoid inadvertent electrode placement.
Answers are grounded in this paper and cite their sources.
Intraoperative lead verification may avert epicardial misplacement; this case leaves open its incidence and systematic prevention.
A 70-year-old woman had an epicardial pacing system implanted after transvenous right ventricular endocardial pacing resulted in cardiac perforation. At the time of revision of the epicardial system for pacing failure, it was noted that one of the bipolar epicardial electrodes was pacing atrial tissue. The differential diagnosis of this unusual finding is discussed and recommendation urged for systematic analysis of pacing system function at the time of implant, in order to avoid inadvertent electrode placement.
No takes yet. Share an insight, caveat, or question.
Lok et al. (1981) conducted a case report in Pacing failure (n=1). Epicardial pacing system implantation was evaluated on Inadvertent atrial location of epicardial pacemaker electrode. Revision of an epicardial pacing system for pacing failure in a 70-year-old woman revealed inadvertent placement of a bipolar epicardial electrode on atrial tissue.