Key result
A smartphone-based blood pressure monitor using PPG technology achieved >90% average accuracy and strong correlation for predicting reference BP values.
Why the study?
The study aimed to develop a user-friendly mobile personal health care system for elderly people to enable noninvasive, pervasive, and continuous estimation of blood pressure level and variability.
Does a wearable PPG-only sensor and smartphone app accurately estimate blood pressure compared to reference values?
Does a wearable PPG-only sensor and smartphone app accurately estimate blood pressure compared to reference values?
A novel PPG-based smartphone system showed high correlation for BP estimation, but only diastolic BP met clinical accuracy standards, indicating further development is needed before clinical use.
Answers are grounded in this paper and cite their sources.
May enable accessible BP monitoring; leaves open need for validation against full clinical accuracy standards before adoption.
BACKGROUND Smartphone-based blood pressure (BP) monitoring using photoplethysmography (PPG) technology has emerged as a promising approach to empower users with self-monitoring for effective diagnosis and control of hypertension. OBJECTIVE This study aimed to develop a mobile personal health care system for noninvasive, pervasive, and continuous estimation of BP level and variability, which is user friendly for elderly people. METHODS The proposed approach was integrated by a self-designed cuffless, calibration-free, wireless, and wearable PPG-only sensor and a native purposely designed smartphone app using multilayer perceptron machine learning techniques from raw signals. We performed a development and usability study with three older adults (mean age 61.3 years, SD 1.5 years; 66% women) to test the usability and accuracy of the smartphone-based BP monitor. RESULTS The employed artificial neural network model had good average accuracy (>90%) and very strong correlation (>0.90) (P<.001) for predicting the reference BP values of our validation sample (n=150). Bland-Altman plots showed that most of the errors for BP prediction were less than 10 mmHg. However, according to the Association for the Advancement of Medical Instrumentation and British Hypertension Society standards, only diastolic blood pressure prediction met the clinically accepted accuracy thresholds. CONCLUSIONS With further development and validation, the proposed system could provide a cost-effective strategy to improve the quality and coverage of health care, particularly in rural zones, areas lacking physicians, and areas with solitary elderly populations.
No takes yet. Share an insight, caveat, or question.
Mena et al. (2020) studied Hypertension (n=3). Smartphone-based blood pressure monitor using photoplethysmography (PPG) vs. Reference BP values was evaluated on Accuracy and correlation for predicting reference BP values (Correlation >0.90, p=<.001). A smartphone-based blood pressure monitor using PPG technology achieved >90% average accuracy and strong correlation for predicting reference BP values.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: