Validation study demonstrates accurate myocardial T1 mapping at 5T using real-time spoiled gradient echo in healthy volunteers, suggesting improved precision over standard MOLLI methods.
Objective. To develop an accurate myocardial T 1 mapping technique at 5T using Look-Locker-based multiple inversion-recovery with the real-time spoiled gradient echo (GRE) acquisition. Approach. The proposed T 1 mapping technique (mIR-rt) samples the recovery of inverted magnetization using the real-time GRE and the images captured during diastole are selected for T 1 fitting. Multiple-inversion recoveries are employed to increase the sample size for accurate fitting. The T 1 mapping method was validated using Bloch simulation, phantom studies, and in 16 healthy volunteers at 5T. Main Results. In both simulation and phantom studies, the T 1 values measured by mIR-rt closely approximate the reference T 1 values, with errors less than 3%, while the conventional MOLLI sequence underestimates T 1 values. The myocardial T 1 values at 5T are 1553 ± 52 ms, 1531 ± 53 ms, and 1526 ± 60 ms (mean ± standard deviation) at the apex, middle, and base, respectively. The T 1 values measured by MOLLI (1350 ± 48 ms, 1349 ± 47 ms, and 1354 ± 45 ms at the apex, middle, and base) were significantly lower than those of mIR-rt with p < 0.001 for all three layers. The mIR-rt sequence method used in our study provides high reproducibility, particularly in the middle slices, supporting its practical relevance for myocardial T 1 mapping. Significance. The proposed method is feasible for myocardial T 1 mapping at 5T and provides better accuracy than the conventional MOLLI sequence.
No takes yet. Share an insight, caveat, or question.
Ge et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: