Gradient acoustic noise has been measured and characterized for an epoxy-potted, shielded gradient assembly in a 1.5 T MRI system. Noise levels vary by 10 dB or more as a function of longitudinal position in the scanner and reflect the pattern of forces applied to the gradient assembly. The noise level increases slightly (1-3 dB) with a patient in the scanner. The spectrum of the noise is similar (but not identical) to the spectrum of the input signal. A gradient-pulse-to-acoustic-noise transfer function was obtained by using a white noise voltage input to the gradient system. The transfer function enabled us to accurately predict acoustic noise output for a pulse sequence consisting of a series of trapezoidal pulses on a single axis and for a clinical fast spin echo sequence with gradients present on all three axes.