Key result
A pediatric circulatory computer model matched human hemodynamics with 95% accuracy, predicting that pulsatile VADs and IABPs restore aortic pulsatility while continuous flow VADs diminish it.
Why the study?
How do simulated continuous flow VADs, pulsatile flow VADs, and IABPs affect acute hemodynamic responses in a computer model of pediatric left ventricular failure?
How do simulated continuous flow VADs, pulsatile flow VADs, and IABPs affect acute hemodynamic responses in a computer model of pediatric left ventricular failure?
A newly developed computer model of the pediatric circulatory system accurately simulates hemodynamics and predicts that pulsatile VADs and IABPs restore pulsatility better than continuous flow VADs in pediatric heart failure.
Pediatric model predictions should not guide device choice; hypothesis-generating for pulsatility effects and requires clinical validation.
Lumped parameter computer models of the pediatric circulatory systems for 1- and 4-year-olds were developed to predict hemodynamic responses to mechanical circulatory support devices. Model parameters, including resistance, compliance and volume, were adjusted to match hemodynamic pressure and flow waveforms, pressure-volume loops, percent systole, and heart rate of pediatric patients (n = 6) with normal ventricles. Left ventricular failure was modeled by adjusting the time-varying compliance curve of the left heart to produce aortic pressures and cardiac outputs consistent with those observed clinically. Models of pediatric continuous flow (CF) and pulsatile flow (PF) ventricular assist devices (VAD) and intraaortic balloon pump (IABP) were developed and integrated into the heart failure pediatric circulatory system models. Computer simulations were conducted to predict acute hemodynamic responses to PF and CF VAD operating at 50%, 75% and 100% support and 2.5 and 5 ml IABP operating at 1:1 and 1:2 support modes. The computer model of the pediatric circulation matched the human pediatric hemodynamic waveform morphology to within 90% and cardiac function parameters with 95% accuracy. The computer model predicted PF VAD and IABP restore aortic pressure pulsatility and variation in end-systolic and end-diastolic volume, but diminish with increasing CF VAD support.
No takes yet. Share an insight, caveat, or question.
Giridharan et al. (2007) studied Pediatric heart failure (n=6). Simulated continuous flow (CF) and pulsatile flow (PF) VAD and IABP was evaluated on Hemodynamic waveform morphology and cardiac function parameters. A pediatric circulatory computer model matched human hemodynamics with 95% accuracy, predicting that pulsatile VADs and IABPs restore aortic pulsatility while continuous flow VADs diminish it.