Key result
Catheters with additional high precision bipoles detected CTI conduction gaps that correlated perfectly with incomplete functional block diagnosed by incremental pacing (Spearman's rho=1, P<0.01).
Why the study?
Do catheters equipped with additional high precision bipoles (AHPB) accurately detect conduction gaps compared to the incremental pacing maneuver during CTI ablation for atrial flutter?
Observational (n=20)
Do catheters equipped with additional high precision bipoles (AHPB) accurately detect conduction gaps compared to the incremental pacing maneuver during CTI ablation for atrial flutter?
Answers are grounded in this paper and cite their sources.
Catheters equipped with additional high precision bipoles can effectively detect conduction gaps during CTI ablation, correlating perfectly with the incremental pacing maneuver.
INTRODUCTION: The incremental pacing (IP) maneuver is a highly specific technique that improves the ability to confirm complete CTI conduction block during typical atrial flutter (AFL) ablation, and reduces long-term AFL recurrences. The purpose of this study is to assess the performance of new catheters equipped with additional high precision bipoles (AHPB) to allow the visualization of the cavotricuspid isthmus (CTI) conduction gap and to compare them with the IP maneuver. METHODS AND RESULTS: Twenty consecutive patients undergoing catheter ablation of the CTI for AFL were included. The IP maneuver confirmed functional versus complete CTI block. Local electrogram analysis using AHPB was then used to assess the presence or absence of gaps across the CTI line. Mean age was 67 years and 80% were male. At the end of the procedure CTI block was achieved in all patients. A transient stage of functional CTI block was observed in 40%. In all cases a continuous fragmented electrogram was present between the double potentials in the CTI in the AHPB channels. In contrast, no electrogram was observed between the CTI double potentials in any of the 20 patients once complete block was confirmed by the IP maneuver. When both techniques were compared a significant association and correlation were observed (chi-square <0.01, Spearman's rho = 1, P < 0.01). CONCLUSION: Catheters equipped with AHPB can aid in the assessment of complete CTI block during AFL ablation procedures by detecting conduction gaps that correlate with incomplete functional block diagnosed by the IP maneuver.
No takes yet. Share an insight, caveat, or question.
Vallès et al. (2016) conducted an observational in Typical atrial flutter (AFL) (n=20). Catheters equipped with additional high precision bipoles (AHPB) vs. Incremental pacing (IP) maneuver was evaluated on Correlation between AHPB gap detection and IP maneuver for CTI block assessment (Spearman's rho = 1, p=< 0.01). Catheters with additional high precision bipoles detected CTI conduction gaps that correlated perfectly with incomplete functional block diagnosed by incremental pacing (Spearman's rho=1, P<0.01).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: