Key result
Higher nocturnal respiratory rate on continuous mattress monitoring strongly predicts 30-day heart failure readmission.
Why the study?
Does continuous nocturnal physiological monitoring using an under-the-mattress piezoelectric sensor predict hospital readmission in patients recently discharged for heart failure?
Observational (n=30)
No
Does continuous nocturnal physiological monitoring using an under-the-mattress piezoelectric sensor predict hospital readmission in patients recently discharged for heart failure?
Continuous nocturnal monitoring with a contactless piezoelectric sensor is feasible and identifies elevated respiratory rate as a predictor of heart failure readmission.
Answers are grounded in this paper and cite their sources.
May aid post-discharge HF risk stratification; leaves open whether respiratory rate–guided interventions reduce readmissions.
Introduction A piezoelectric sensor (PS) converts mechanical deformations into electrical signals. We used a contactless under-the-mattress PS to monitor physiological vibrations resulting from breathing, pumping of the heart, and body movements, among individuals at home following hospitalization for heart failure (HF). Our objectives were to assess acceptability of the device in the home, to assess physiological patterns, and to determine if altered patterns correlate with readmission. Methods We conducted a prospective observational study of 30 patients discharged home following HF hospitalization. PS data included a continuous nightly assessment of heart rate, respiration rate, movement rate, rapid and shallow respiration duration, and a behaviour score. We utilized random forest classification to classify average nightly data by readmission status. Results We collected 640 nights of PS data from 29 patients. There were nine readmissions, of which four were for HF. PS monitoring was tolerated by all but one of the participants. We inspected continuous nightly physiological profiles and noted differences between patients who were and were not readmitted. Patients readmitted for HF had higher average heart and respiration rates, and more respiration variability. Average nightly respiratory rate was most predictive of readmission. Discussion We are the first to study nocturnal physiological patterns of HF patients at home using a contactless under-the-mattress monitoring system. We noted patterns that may be unique to patients at risk for readmission due to HF. Respiratory rate was the most important risk-adjusted associate of readmission for HF. Further studies should investigate the efficacy of home PS monitoring in HF populations.
No takes yet. Share an insight, caveat, or question.
Bennett et al. (2015) conducted an observational in Heart Failure (n=30). Under-the-mattress piezoelectric sensor (EverOn) was evaluated on Hospital readmission within 30 days of hospital discharge. Continuous nocturnal monitoring with an under-the-mattress piezoelectric sensor identified that higher average nightly respiratory rate was the most important predictor of 30-day heart failure readmission.