Why the study?
Head-to-head comparisons of ECG monitoring between adhesive single-lead and Holter devices were limited for arrhythmias other than AF.
Does an adhesive single-lead ECG device provide comparable diagnostic accuracy to Holter monitoring in non-AF patients with general arrhythmia?
Does an adhesive single-lead ECG device provide comparable diagnostic accuracy to Holter monitoring in non-AF patients with general arrhythmia?
An adhesive single-lead ECG device provides comparable diagnostic accuracy and ECG parameter measurement to standard Holter monitoring for 24-hour ambulatory monitoring in non-AF patients.
Supports adhesive single-lead ECG as potential Holter alternative in non-AF arrhythmia monitoring; hypothesis-generating, requiring prospective validation before practice change.
There are few reports on head-to-head comparisons of electrocardiogram (ECG) monitoring between adhesive single-lead and Holter devices for arrhythmias other than atrial fibrillation (AF). This study aimed to compare 24 h ECG monitoring between the two devices in patients with general arrhythmia. Twenty-nine non-AF patients with a workup of pre-diagnosed arrhythmias or suspicious arrhythmic episodes were evaluated. Each participant wore both devices simultaneously, and the cardiac rhythm was monitored for 24 h. Selective ECG parameters were compared between the two devices. Two cardiologists independently compared the diagnoses of each device. The two most frequent monitoring indications were workup of premature atrial contractions (41.4%) and suspicious arrhythmia-related symptoms (37.9%). The single-lead device had a higher noise burden than the Holter device (0.04 ± 0.05% vs. 0.01 ± 0.01%, p = 0.024). The number of total QRS complexes, ventricular ectopic beats, and supraventricular ectopic beats showed an excellent degree of agreement between the two devices (intraclass correlation coefficients = 0.991, 1.000, and 0.987, respectively). In addition, the minimum/average/maximum heart rates showed an excellent degree of agreement. The two cardiologists made coherent diagnoses for all 29 participants using both monitoring methods. In conclusion, the single-lead adhesive device could be an acceptable alternative for ambulatory ECG monitoring in patients with general arrhythmia.
No takes yet. Share an insight, caveat, or question.
Kwon et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: