Doppler echocardiographic assessment of the St. Jude Medical prosthetic valve in the aortic position using the continuity equation.

View Full Paper
ECEdward ChafizadehWZWilliam A. Zoghbi

Structured PICO

Does Doppler echocardiographic assessment using the continuity equation improve the evaluation of St. Jude Medical prosthetic aortic valves compared to using gradients alone?

P
Population
67 clinically stable patients (mean age 58 +/- 14 years) without evidence of valve dysfunction, evaluated within 10 +/- 6 days after aortic valve replacement with St. Jude Medical valves (valve size 19 to 31 mm, ejection fraction 30% to 75%).
I
Intervention
Doppler echocardiographic assessment using the continuity equation to calculate effective prosthetic aortic valve area.
C
Comparator
Assessment using Doppler-derived maximal gradients alone.
O
Outcome
Correlation of effective prosthetic aortic valve area with actual valve orifice area and differentiation of valve sizes.surrogate

The continuity equation provides an improved assessment of St. Jude prosthetic aortic valve function compared to using gradients alone by accounting for flow through the valve.

Abstract

To test whether the continuity equation can be applied to the noninvasive assessment of prosthetic aortic valve function, Doppler echocardiography was performed in 67 patients (mean age, 58 +/- 14 years) within 10 +/- 6 days after valve replacement with St. Jude Medical valves. All patients were clinically stable and without evidence of valve dysfunction. Valve size ranged from 19 to 31 mm, and ejection fraction ranged from 30% to 75%. With the parasternal long-axis view, the left ventricular outflow diameter measured just proximal to the prosthetic valve correlated well with valve size (r = 0.92). Doppler-derived maximal gradients ranged from 9 to 71 mm Hg. Effective prosthetic aortic valve area by the continuity equation ranged between 0.73 cm2 for a 19-mm valve and 4.23 cm2 for a 31-mm valve. With analysis of variance, effective orifice area differentiated various valve sizes (p less than 10(-14)) better than did gradients alone (p = 0.003) and correlated better with actual valve orifice area (r = 0.83 versus - 0.40). A Doppler velocity index, the ratio of peak velocity in the left ventricular outflow to that of the aortic jet, averaged 0.41 +/- 0.09 and was less dependent on valve size (r = 0.43). Thus, the continuity equation can be applied to the assessment of prosthetic St. Jude valves in the aortic position. By accounting for flow through the valve, it provides an improved assessment over the sole use of gradients in the evaluation of prosthetic valve function.