Key result
Active-fixation pacing leads demonstrated stable electrical performance comparable to passive-fixation leads over a five-year follow-up, with no difference in pacing thresholds at 5 years.
Why the study?
Does active-fixation pacing lead use improve long-term electrical performance compared to passive-fixation leads in newly implanted patients?
RCT (n=199)
Random number generated by computer
No
Does active-fixation pacing lead use improve long-term electrical performance compared to passive-fixation leads in newly implanted patients?
Active-fixation pacing leads demonstrate stable long-term electrical performance that is comparable to passive-fixation leads over a 5-year follow-up.
Answers are grounded in this paper and cite their sources.
Active-fixation leads showed comparable 5-year thresholds in this cohort; leaves open whether randomized data would confirm equivalence for lead selection.
Active-fixation pacing leads allow the use of selective pacing sites. We evaluated their long-term performance versus passive-fixation leads in 199 newly implanted patients (n = 100 active and n = 99 passive). Postoperative pacing thresholds in the active group were higher than in the passive group (0.85 ± 0.31 V vs. 0.53 ± 0.21 V at baseline, P < 0.001). The active thresholds fell to 0.72 ± 0.23 V at 5 years with a significant drop at one month (0.68 ± 0.53 V, P = 0.003). The passive thresholds slightly increased to 0.72 ± 0.31 V at five years. Differences between groups were significant until three years (all P < 0.05). Active impedances were generally lower than passive impedances (600.44 ± 94.31Ω vs. 683.14 ± 110.98Ω at baseline), and both showed significant reductions at one month to 537.96 ± 147.43Ω in the active group, and after three months to 643.85 ± 82.40Ω in the passive group (both P < 0.01 vs. baseline). Impedance differences between groups were significant until four years (all P < 0.05). Adverse events included thresholds over 1 V, 5 of 6 active and 2 of 5 passive leads returned to below 1 V. One active left ventricular lead dislodged. One passive left subclavian lead insulation fracture occurred. Thus Active fixation pacing leads are stable in a five-year long-term follow up. There was no difference between active and passive leads in terms of electrical performance.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2015) conducted an RCT in Bradyarrhythmia requiring permanent pacemaker implantation (n=199). Active-fixation pacing leads vs. Passive-fixation pacing leads was evaluated on Pacing threshold at 5 years (V). Active-fixation pacing leads demonstrated stable electrical performance comparable to passive-fixation leads over a five-year follow-up, with no difference in pacing thresholds at 5 years.