Key result
Medtronic 4058 and 4058M bipolar atrial screw-in leads demonstrated a 93.5% 1-year complication-free survival rate, with a tendency toward high pacing thresholds at 3 months.
Why the study?
What are the short- and long-term complication rates and performance characteristics of the Medtronic 4058/4058M bipolar atrial screw-in lead?
Observational (n=136)
No
What are the short- and long-term complication rates and performance characteristics of the Medtronic 4058/4058M bipolar atrial screw-in lead?
Answers are grounded in this paper and cite their sources.
The Medtronic 4058/4058M atrial pacing lead demonstrates favorable acute and chronic performance with 93.5% 1-year complication-free survival, but exhibits a tendency toward high pacing thresholds at 3 months.
Since 1989, 136 Medtronic 4058 and 4058M bipolar atrial screw-in leads have been implanted at the Mayo Clinic. Early lead related complications included dislodgment in 4 (2.9%). Over a median follow-up time of 14.4 months (1 day to 3.3 years), there were 11 lead related complications (undersensing, failure to capture, diaphragmatic pacing, and gross lead dislodgment). Chronic complications resulted in reoperations in four patients (2.9%). Of 77 patients in whom pacing thresholds were measured between 2 and 4 months after implantation, 9 (11.7%) and 2 (2.6%) had high pacing thresholds and very high thresholds, respectively. The Kaplan-Meier estimate of the probability of 1-year complication-free lead survival was 93.5%. There were no lead material failures. We conclude that the 4058/4058M lead implanted in the atrial position has favorable acute and chronic performance data, with a tendency toward high pacing thresholds at 3 months. The cause of this phenomenon and its course over time should be further evaluated.
No takes yet. Share an insight, caveat, or question.
Glikson et al. (1996) conducted an observational in Atrial pacing (n=136). Medtronic 4058 and 4058M bipolar atrial screw-in leads was evaluated on 1-year complication-free lead survival. Medtronic 4058 and 4058M bipolar atrial screw-in leads demonstrated a 93.5% 1-year complication-free survival rate, with a tendency toward high pacing thresholds at 3 months.