Key result
Polyurethane atrial leads demonstrated an 88% cumulative survival rate at 5 years, with only 6% experiencing pacing or sensing failures.
Why the study?
Do polyurethane atrial leads provide reliable long-term pacing and sensing function in patients requiring pacemakers?
Observational (n=494)
No
Do polyurethane atrial leads provide reliable long-term pacing and sensing function in patients requiring pacemakers?
Answers are grounded in this paper and cite their sources.
Polyurethane atrial leads demonstrate reliable long-term functional integrity with 88% survival at 5 years, supporting their use based on clinical indication rather than concerns of lead unreliability.
The effectiveness and reliability of atrial leads has been questioned. We studied retrospectively, all atrial leads implanted at our center (n = 494; 438 Medtronic Model 6957J, 56 Medtronic Model 4512) over a 5-year period ending December 31, 1987, to determine the frequency of atrial lead failure (pacing, sensing, or both) and the median duration of proper pacing and sensing function for each lead model studied. Eighty-eight percent of the polyurethane atrial leads continued to function satisfactorily at 5 years, results somewhat better than those reported heretofore in the literature, as well as our own past results with a variety of different lead types. There were 29 failures of pacing, sensing, or both (6% of implants). The cumulative survival of the atrial leads at 5 years was 88%. Pacing and sensing survival were 91% +/- 2.4% and 88% +/- 2.9%, respectively. We conclude that the choice of pacing mode for a new pacemaker should be based solely on the clinical indication and not on the concern that atrial pacing and sensing will be unreliable.
No takes yet. Share an insight, caveat, or question.
Parsonnet et al. (1991) conducted an observational in Pacemaker implantation (n=494). Polyurethane atrial leads (Medtronic Model 6957J and 4512) was evaluated on Cumulative survival of atrial leads (freedom from pacing or sensing failure) at 5 years. Polyurethane atrial leads demonstrated an 88% cumulative survival rate at 5 years, with only 6% experiencing pacing or sensing failures.
Synapse has enriched 10 related papers. Consider them for comparative context: