Key result
Multifilar silicone rubber insulated pacemaker leads demonstrated superior longevity compared to monofilar leads, achieving a 15-year cumulative survival rate of 99% to 100%.
Why the study?
Do different pacemaker lead materials and designs affect long-term lead survival?
Population
5,405 pacemaker leads implanted at Montefiore Medical Center between 1960 and May 31, 1993
Comparison
Multifilar silicone rubber insulated pacemaker… vs Monofilar silicone rubber leads and other…
Design
Cohort
Follow-up
up to 24 years
Authors
Loading...
Multifilar silicone leads may improve pacemaker longevity; leaves open need for randomized confirmation in modern devices.
Cohort (n=5,405)
No
Do different pacemaker lead materials and designs affect long-term lead survival?
Multifilar silicone rubber insulated pacemaker leads demonstrate superior long-term survival compared to monofilar leads, providing historical benchmarks for lead quality and longevity.
Furman et al. (1995) conducted a cohort in Pacemaker implantation (n=5,405). Multifilar silicone rubber insulated leads vs. Monofilar silicone rubber leads was evaluated on Cumulative survival rate (CSR). Multifilar silicone rubber insulated pacemaker leads demonstrated superior longevity compared to monofilar leads, achieving a 15-year cumulative survival rate of 99% to 100%.