Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
January 1, 1994Pacing and Clinical Electrophysiology

Long‐Term Performance of Endocardial Pacing Leads

View Full Paper
Ask AI
Bookmark
Share

Key result

Endocardial pacing leads in older patients (≥65 years) and in the atrial position had fewer verified structural failures (P=0.014 and P=0.007, respectively).

Why the study?

What is the long-term performance and what factors are associated with structural failure of endocardial pacing leads?

Population

2,611 endocardial pacing leads implanted between 1980 and 1991 with at least 1 month of follow-up

Design

Cohort

Follow-up

up to 10 years (minimum 1 month)

Authors

MHMarcelo E. HelgueraRush University Medical CenterJMJames D. MaloneyAmerican College of CardiologySPSergio L. PinskiRush University Medical Center

Discussion

Loading...

Member takes

Implication

May inform age- and position-specific lead selection; extends historical survival data but remains hypothesis-generating.

Study Design

Type

Cohort

Structured PICO

What is the long-term performance and what factors are associated with structural failure of endocardial pacing leads?

P
Population
Medical records of 2,611 endocardial pacing leads implanted between 1980 and 1991 with at least 1 month of follow-up were reviewed to assess long-term performance.
E
Exposure
Endocardial pacing leads
O
Outcome
Structural lead failure (suspected and verified)safety

Endocardial pacing leads demonstrate excellent long-term survival, though performance varies by patient age, implant position, and specific lead model.

Cite This Study

Helguera et al. (1994) conducted a cohort in Endocardial pacing lead structural failure. Endocardial pacing leads was evaluated on Structural lead failure (suspected and verified). Endocardial pacing leads in older patients (≥65 years) and in the atrial position had fewer verified structural failures (P=0.014 and P=0.007, respectively).

synapsesocial.com/papers/6a6a8cbd01245da06ecb38a2https://doi.org/10.1111/j.1540-8159.1994.tb01351.x
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Long‐Term Functional Integrity of Atrial Leads1991 · 27 citations
  2. 2Survival of Implantable Pacemaker Leads1990 · 37 citations
  3. 3A Multicenter Experience with a Bipolar Tined Polyurethane Ventricular Lead1992 · 58 citations
  4. 4Appraisal of Pacing Lead Performance Using Transtelephonic Follow‐Up Data1990 · 9 citations
  5. 5Simultaneous Insulation Deterioration Associated with Side‐by‐Side Subclavian Placement of Two Polyurethane Leads1988 · 45 citations