Key result
The EMT 588 endocardial pacing lead with a small electrode surface area demonstrated a 9-year cumulative survival rate of 84%, compared to 72% for the large electrode series.
Why the study?
What is the long-term durability and nature of failure of the EMT 588 endocardial pacing lead?
Observational (n=1,063)
What is the long-term durability and nature of failure of the EMT 588 endocardial pacing lead?
The EMT 588 endocardial pacing lead showed improved survival with a smaller electrode surface area, though polyethylene insulation remained a vulnerability leading to late failures.
Answers are grounded in this paper and cite their sources.
Smaller electrode surface area may enhance endocardial lead durability; leaves open insulation-related late failure risks in modern devices.
A report is presented of 1253 EMT 588 endocardial leads implanted in 1063 patients. The electrode surface area was large (47 mm2) in the 473 leads implanted during 1962-1973, and small (8-12 mm2) in the 780 leads implanted in 1974-1981. Replacement of 245 leads was necessary, in 187 cases due to lead failure and in 58 because of problems with normally functioning leads, such as infection. The highest lead failure rate occurred within the first 6 months after implantation, and was mostly caused by displacement and threshold increase. The dominating cause of late failure was lead lesion. The cumulative lead survival rates for the first and the second series were 88 and 94%, respectively after 1 year, 80 and 91% after 5 years, and 72 and 84% after 9 years. The polyethylene insulation proved to be the most vulnerable part of this lead. Insulation lesions could result in corrosion and/or fracture of the steel conductors. The chronic thresholds were largely stable.
No takes yet. Share an insight, caveat, or question.
MAJ LEVANDER‐LINDGREN (1988) conducted an observational in Patients requiring endocardial pacing (n=1,063). EMT 588 endocardial pacing lead (small electrode surface area, 8-12 mm2) vs. EMT 588 endocardial pacing lead (large electrode surface area, 47 mm2) was evaluated on Cumulative lead survival rate at 9 years. The EMT 588 endocardial pacing lead with a small electrode surface area demonstrated a 9-year cumulative survival rate of 84%, compared to 72% for the large electrode series.
Synapse has enriched 10 related papers. Consider them for comparative context: