July 1, 1992Pacing and Clinical Electrophysiology

A Multicenter Experience with a Bipolar Tined Polyurethane Ventricular Lead

View Full Paper
Ask AI
Bookmark
Share

Key result

The Medtronic 4012 pacing lead demonstrated an unacceptable failure rate, reaching a 20.9% probability of failure at 6 years after implantation.

Why the study?

What is the failure rate and clinical manifestation of failure for the Medtronic 4012 polyurethane bipolar tined pacing lead?

Population

1,190 Medtronic 4012 pacing leads implanted in patients across 6 centers in the United States and Canada.

Design

Cohort

Follow-up

6 years

Authors

David L. HayesDavid L. HayesElectrophysiologyKGKevin J. GrahamNortheastern Illinois Regional Crime LaboratoryMIMarleen IrwinUniversity of Southern California

Discussion

Loading...

Member takes

Implication

May warrant enhanced surveillance for Medtronic 4012 leads; leaves open need for prospective validation of failure rates.

Study Design

Type

Cohort (n=1,190)

Multicenter

Yes

Structured PICO

What is the failure rate and clinical manifestation of failure for the Medtronic 4012 polyurethane bipolar tined pacing lead?

P
Population
1,190 Medtronic 4012 pacing leads implanted across six centers in the United States and Canada, evaluated for lead failure over 6 years.
E
Exposure
Medtronic 4012 pacing lead (polyurethane bipolar tined ventricular lead)
O
Outcome
Probability and clinical manifestations of lead failure (specifically failures compatible with an insulation problem)safety

The Medtronic 4012 pacing lead has an unacceptably high failure rate of 20.9% at 6 years due to insulation problems, prompting recommendations for prophylactic replacement in pacemaker-dependent patients.

Cite This Study

Hayes et al. (1992) conducted a cohort in Patients requiring ventricular pacing (n=1,190). Medtronic 4012 pacing lead was evaluated on Probability of lead failure compatible with an insulation problem. The Medtronic 4012 pacing lead demonstrated an unacceptable failure rate, reaching a 20.9% probability of failure at 6 years after implantation.

synapsesocial.com/papers/6a6aa7d7285b5e3c3db39010https://doi.org/10.1111/j.1540-8159.1992.tb03097.x