Key result
The Medtronic 4012 pacing lead demonstrated an unacceptable failure rate, reaching a 20.9% probability of failure at 6 years after implantation.
Why the study?
What is the failure rate and clinical manifestation of failure for the Medtronic 4012 polyurethane bipolar tined pacing lead?
Population
1,190 Medtronic 4012 pacing leads implanted in patients across 6 centers in the United States and Canada.
Design
Cohort
Follow-up
6 years
Authors
Loading...
May warrant enhanced surveillance for Medtronic 4012 leads; leaves open need for prospective validation of failure rates.
Cohort (n=1,190)
Yes
What is the failure rate and clinical manifestation of failure for the Medtronic 4012 polyurethane bipolar tined pacing lead?
The Medtronic 4012 pacing lead has an unacceptably high failure rate of 20.9% at 6 years due to insulation problems, prompting recommendations for prophylactic replacement in pacemaker-dependent patients.
Hayes et al. (1992) conducted a cohort in Patients requiring ventricular pacing (n=1,190). Medtronic 4012 pacing lead was evaluated on Probability of lead failure compatible with an insulation problem. The Medtronic 4012 pacing lead demonstrated an unacceptable failure rate, reaching a 20.9% probability of failure at 6 years after implantation.