We present a fluorescence lifetime imaging technique with simultaneous spectral and temporal resolution. The technique is fully compatible with the commonly used multiphoton microscopes and nondescanned (direct) detection. An image of the back-aperture of the microscope lens is projected on the input of a fiber bundle. The input of the fiber bundle is circular, and the output is flattened to match the input slit of a spectrograph. The spectrum at the output of the spectrograph is projected on a 16-anode PMT module. For each detected photon, the encoding logics of the PMT module deliver a timing pulse and the number of the PMT channel in which the photon was detected. The photons are accumulated by a multidimensional time-correlated single photon counting (TCSPC) process. The recording process builds up a four-dimensional photon distribution over the times of the photons in the excitation pulse period, the wavelengths of the photons, and the coordinates of the scan area. The method delivers a near-ideal counting efficiency and is capable of resolving double-exponential decay functions. We demonstrate the performance of the technique for autofluorescence imaging of tissue.