Key result
ETPC algorithm achieves ~36% greater energy savings than conventional methods during wearable ECG monitoring.
A novel transmission power control-based algorithm for wearable ECG monitoring systems improves energy efficiency by 35.5% compared to conventional methods.
Answers are grounded in this paper and cite their sources.
Could extend wearable ECG battery life; hypothesis-generating pending clinical validation.
Rapid progress and emerging trends in miniaturized medical devices have enabled the un-obtrusive monitoring of physiological signals and daily activities of everyone's life in a prominent and pervasive manner. Due to the power-constrained nature of conventional wearable sensor devices during ubiquitous sensing (US), energy-efficiency has become one of the highly demanding and debatable issues in healthcare. This paper develops a single chip-based wearable wireless electrocardiogram (ECG) monitoring system by adopting analog front end (AFE) chip model ADS1292R from Texas Instruments. The developed chip collects real-time ECG data with two adopted channels for continuous monitoring of human heart activity. Then, these two channels and the AFE are built into a right leg drive right leg drive (RLD) driver circuit with lead-off detection and medical graded test signal. Human ECG data was collected at 60 beats per minute (BPM) to 120 BPM with 60 Hz noise and considered throughout the experimental set-up. Moreover, notch filter (cutoff frequency 60 Hz), high-pass filter (cutoff frequency 0.67 Hz), and low-pass filter (cutoff frequency 100 Hz) with cut-off frequencies of 60 Hz, 0.67 Hz, and 100 Hz, respectively, were designed with bilinear transformation for rectifying the power-line noise and artifacts while extracting real-time ECG signals. Finally, a transmission power control-based energy-efficient (ETPC) algorithm is proposed, implemented on the hardware and then compared with the several conventional TPC methods. Experimental results reveal that our developed chip collects real-time ECG data efficiently, and the proposed ETPC algorithm achieves higher energy savings of 35.5% with a slightly larger packet loss ratio (PLR) as compared to conventional TPC (e.g., constant TPC, Gao's, and Xiao's methods).
No takes yet. Share an insight, caveat, or question.
Sodhro et al. (2018) studied Healthy (no history of cardiovascular disease) (n=30). Energy-efficient transmission power control (ETPC) algorithm vs. Conventional TPC methods (constant TPC, Gao's, and Xiao's methods) was evaluated on Energy savings. The proposed ETPC algorithm achieved 35.5% higher energy savings compared to conventional transmission power control methods during real-time ECG data collection.
Synapse has enriched 10 related papers. Consider them for comparative context: