Key result
A wearable ECG processor using a short-term autocorrelation algorithm for robust instantaneous heart rate monitoring consumed 13.7 μA for heart rate logging applications.
The development of a low-power (13.7 μA) noise-tolerant ECG processor enables more efficient wearable bio-signal monitoring systems for daily life.
May enable extended wearable HR monitoring; leaves open clinical validation studies.
To prevent lifestyle diseases, wearable bio-signal monitoring systems for daily life monitoring have attracted attention. Wearable systems have strict size and weight constraints, which impose significant limitations of the battery capacity and the signal-to-noise ratio of bio-signals. This report describes an electrocardiograph (ECG) processor for use with a wearable healthcare system. It comprises an analog front end, a 12-bit ADC, a robust Instantaneous Heart Rate (IHR) monitor, a 32-bit Cortex-M0 core, and 64 Kbyte Ferroelectric Random Access Memory (FeRAM). The IHR monitor uses a short-term autocorrelation (STAC) algorithm to improve the heart-rate detection accuracy despite its use in noisy conditions. The ECG processor chip consumes 13.7 μA for heart rate logging application.
No takes yet. Share an insight, caveat, or question.
Izumi et al. (2014) studied this question. Wearable ECG processor with STAC algorithm was evaluated on Power consumption and heart-rate detection accuracy. A wearable ECG processor using a short-term autocorrelation algorithm for robust instantaneous heart rate monitoring consumed 13.7 μA for heart rate logging applications.