Key result
An ultra-low power ECG-on-chip was developed with a core area of 5.74 mm2 and power consumption of 2.3 μW, demonstrating suitability for long-term cardiac monitoring in wearable devices.
The development of a 2.3 μW ECG-on-chip provides a viable, ultra-low power solution for long-term cardiac monitoring in wearable sensors.
Answers are grounded in this paper and cite their sources.
May enable continuous wearable ECG monitoring; extends low-power designs but leaves clinical validation open.
This brief presents an ultra-low power single chip solution for electrocardiography (ECG) signal acquisition and processing in wearable ECG sensors. The chip contains a low noise preamplifier with embedded band-pass function, a programmable gain buffer, a 12-bit successive approximation ADC, a novel morphological filter based QRS detector, 8-Kb on-chip SRAM, a control unit and MCU interfaces. The chip was designed and implemented in 0.35-μm standard CMOS process. The analog core operates from 0.8 V to 1.8 V, while the digital circuits and SRAM operate from 1.5 V to 3.6 V. The chip has a total core area of 5.74 mm 2 and consumes 2.3 μW. Small size and low power consumption make this chip suitable for usage in wearable ECG sensors. Apart from presenting the measurement results, we also successfully demonstrate a prototype wearable ECG device, for long term cardiac monitoring using the proposed ECG-on-chip.
No takes yet. Share an insight, caveat, or question.
John et al. (2018) studied Cardiac monitoring. ECG-on-chip for wearable sensors was evaluated on Chip performance (power consumption and area). An ultra-low power ECG-on-chip was developed with a core area of 5.74 mm2 and power consumption of 2.3 μW, demonstrating suitability for long-term cardiac monitoring in wearable devices.
Synapse has enriched 10 related papers. Consider them for comparative context: