Key result
Fingertip PPG matches ECG for RMSSD measurement across breathing protocols, but in-ear PPG fails.
Why the study?
Studies assessing respiratory-gated taVNS have been limited to controlled laboratory environments due to reliance on non-portable sensing equipment, creating a need for portable sensor systems for home use.
Do in-ear and fingertip-based PPG sensors accurately measure cardiac vagal tone relevant HRV parameters compared to a gold standard ECG in healthy adults?
Cross-Sectional (n=30)
Randomized order of breathing protocols
No
Do in-ear and fingertip-based PPG sensors accurately measure cardiac vagal tone relevant HRV parameters compared to a gold standard ECG in healthy adults?
Fingertip-based PPG sensors and stretch sensors provide acceptable accuracy for measuring cardiac vagal tone relevant HRV parameters and respiration phases, enabling the development of portable respiratory-gated taVNS systems.
This paper presents a study undertaken to evaluate the sensor systems that were shortlisted to be used in the development of a portable respiratory-gated transcutaneous auricular vagus nerve stimulation (taVNS) system. To date, all published studies assessing respiratory-gated taVNS have been performed in controlled laboratory environments. This limitation arises from the reliance on non-portable sensing equipment, which poses significant logistical challenges. Therefore, we recognised a need to develop a portable sensor system for future research, enabling participants to perform respiratory-gated stimulation conveniently from their homes. This study aimed to measure the accuracy of an in-ear and a fingertip-based photoplethysmography (PPG) sensor in measuring cardiac vagal tone relevant heart rate variability (HRV) parameters of root mean square of successive R-R interval differences (RMSSDs) and the high-frequency (HF) component of HRV. Thirty healthy participants wore the prototype sensor equipment and the gold standard electrocardiogram (ECG) equipment to record beat-to-beat intervals simultaneously during 10 min of normal breathing and 10 min of deep slow breathing (DSB). Additionally, a stretch sensor was evaluated to measure its accuracy in detecting exhalation when compared to the gold standard sensor. We used Bland-Altman analysis to establish the agreement between the prototypes and the ECG system. Intraclass correlation coefficients (ICCs) were calculated to establish consistency between the prototypes and the ECG system. For the stretch sensor, the true positive rate (TPR), false positive rate (FPR), and false negative rate (FNR) were calculated. Results indicate that while ICC values were generally good to excellent, only the fingertip-based sensor had an acceptable level of agreement in measuring RMSSDs during both breathing phases. Only the fingertip-based sensor had an acceptable level of agreement during normal breathing in measuring HF-HRV. The study highlights that a high correlation between sensors does not necessarily translate into a high level of agreement. In the case of the stretch sensor, it had an acceptable level of accuracy with a mean TPR of 85% during normal breathing and 95% during DSB. The results show that the fingertip-based sensor and the stretch sensor had acceptable levels of accuracy for use in the development of the respiratory-gated taVNS system.
No takes yet. Share an insight, caveat, or question.
Parikh et al. (2025) conducted a cross-sectional in Healthy adults (n=30). Fingertip-based and in-ear photoplethysmography (PPG) sensors vs. Gold standard 12-lead ECG system was evaluated on Agreement of RMSSD and HF-HRV measurements between PPG sensors and ECG. The fingertip-based PPG sensor demonstrated an acceptable level of agreement with gold standard ECG for measuring RMSSD during both normal and deep slow breathing, whereas the in-ear sensor did not.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: