April 13, 2021Journal of Medical Engineering & Technology

Both co-primary endpoints for HR and RR resulted in 100% accuracy (95% CI: HR 98.8–100%; RR 98.9–100%).

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Remote patient monitoring is needed for resource limitations and infection control, but current sensors require physical contact that can cause discomfort or non-adherence.

Does a novel contact-free optical system accurately measure heart and respiratory rate compared to a standard bedside monitor?

Population

115 patients who completed study procedures

Comparison

Contact-free optical system (Gili Pro BioSensor) vs standard reference bed-side monitor (Mindray)

Authors

OHOfer HavakukBSBen SadehIMIlan Merdler

Discussion

Loading...

Member takes

Overview

May enable comfortable remote vital-sign monitoring; leaves open larger validation before clinical adoption.

Structured PICO

Does a novel contact-free optical system accurately measure heart and respiratory rate compared to a standard bedside monitor?

P
Population
115 patients, mean age 66 ± 14.6, 60% male, 31% obese (BMI > 30 kg/m2).
I
Intervention
Contact-free optical system measuring nano-vibrations and movements (Gili Pro BioSensor)
C
Comparator
Standard reference bed-side monitor with electrocardiogram and capnograph modules (Mindray)
O
Outcome
HR and RR accuracy (HR in subjects without active arrhythmias, RR in all subjects)surrogate

A novel contact-free optical system demonstrated over 99% accuracy in detecting heart and respiratory rate compared to standard bedside monitors.

Cite This Study

Havakuk et al. (2021) studied this question.

synapsesocial.com/papers/6a7acfc79092edbae3f3bcfahttps://doi.org/10.1080/03091902.2021.1905896
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Contactless monitoring of respiratory rate (RR) and heart rate (HR) in non-acuity settings: a clinical validity study2022 · 8 citations
  2. 2Contactless respiratory and heart rate monitoring: validation of an innovative tool2010 · 53 citations
  3. 3Validation of vital sign measurement by nanosecond pulse near-field sensing and the conductive silver paste printed textile technique2022
  4. 4Validation of a Novel Fiber-Optic Sensor System for Monitoring Cardiorespiratory Activities During MRI Examinations2017 · 26 citations
  5. 5Clinical validation of a contactless respiration rate monitor2023 · 35 citations