Why the study?
Wearable technology for vital sign monitoring faces a tradeoff between accuracy and comfort caused by the elastic textile used.
Does the iSmartweaR wearable sensor accurately measure heart rate and breathing rate compared to gold standard methods in healthy participants and patients with cardiovascular disease?
Does the iSmartweaR wearable sensor accurately measure heart rate and breathing rate compared to gold standard methods in healthy participants and patients with cardiovascular disease?
The iSmartweaR wearable sensor provides accurate, real-time, and continuous monitoring of heart rate and breathing rate across different sexes and static postures.
Answers are grounded in this paper and cite their sources.
Requires further validation before clinical use; leaves open accuracy across postures, sex, and real-world settings.
Vital sign monitoring is fundamental in daily patient care. Wearable technology has become an important tool in personalized medicine; however, there is a tradeoff problem between accuracy and comfort caused by the elastic textile used. Our study sought to examine the accuracy of heart rate (HR) and breathing rate (BR) monitoring as measured by iSmartweaR, a novel wearable and wireless sensor that uses the laser Doppler technique, with special focus on sex and different static postures. A total of 30 male and 23 female patients, including 23 patients with cardiovascular diseases and 30 healthy participants, were recruited. The HR and BR were measured with iSmartweaR and the gold standard method (CardioSoft and MasterScreen™ CPX) simultaneously in three static postures (supine rest, right lateral decubitus, and left lateral decubitus). The mean errors of HR and BR between iSmartweaR and the gold standard method were close to zero (0.55 and 0.71, respectively). The standard deviations of the HR error and BR error were 0.59 and 0.67, respectively. Disease, posture, sex, and measurement time did not cause clinically significant measurement errors. The overall percentage of missing data was 7.01%. The use of iSmartweaR was valid across different sexes and static postures. It is real-time, continuous and manpower-saving, and can be applied in patient care.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: