Key result
Soft elastomeric sensor shows linear force-resistance response up to 0.2 N for ablation catheters.
Why the study?
Sensing in soft robotics has been challenging due to sensor sensitivity to various loading conditions and nonlinear signal responses under extreme loads.
A novel soft elastomeric sensor demonstrates a predictable linear resistance response to uniaxial force, showing potential for providing force feedback during cardiac catheter ablation.
May support force-feedback sensor development for ablation; leaves open in vivo validation and clinical translation.
The field of soft robotics has attracted the interest of the medical community due to the ability of soft elastic materials to traverse the abnormal environment of the human body. However, sensing in soft robotics has been challenging due to the sensitivity of soft sensors to various loading conditions and the nonlinear signal responses that can arise under extreme loads. Ideally, soft sensors should provide a linear response under a specific loading condition and provide a different response for other loading directions. With these specifications in mind, our team created a soft elastomeric sensor designed to provide force feedback during cardiac catheter ablation surgery. Analytical and computational methods were explored to define a relationship between resistance and applied force for a semicircular, liquid metal filled channel in the soft elastomeric sensor. Pouillet's Law is utilized to calculate the resistance based on the change in cross-sectional area resulting from various applied pressures. FEA simulations were created to simulate the deformation of the sensor under various loads. To confirm the validity of these simulations, the elastomer was modeled as a neo-Hookean material and the liquid metal was modeled as an incompressible fluid with negligible shear modulus under uniaxial compression. Results show a linearly proportional relationship between the resistance of the sensor and the application of a uniaxial force. Altering the direction of applied force results in a quadratic relationship between total resistance and the magnitude of force.
No takes yet. Share an insight, caveat, or question.
Rasmussen et al. (2021) studied Cardiac catheter ablation. Soft resistive elastomeric sensor was evaluated on Change in electrical resistance in response to applied force. A soft elastomeric sensor designed for cardiac catheter ablation demonstrated a linearly proportional relationship between electrical resistance and uniaxial force up to 0.2 N in computational models.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: