Answers are grounded in this paper and cite their sources.
The performance of indirect-detection flat-panel imagers incorporating CsI:Tl x-ray converters is examined through calculation of the detective quantum efficiency (DQE) under conditions of chest radiography, fluoroscopy, and mammography. Calculations are based upon a cascaded systems model which has demonstrated excellent agreement with empirical signal, noise- power spectra, and DQE results. For each application, the DQE is calculated as a function of spatial-frequency and CsI:Tl thickness. A preliminary investigation into the optimization of flat-panel imaging systems is described, wherein the x-ray converter thickness which provides optimal DQE for a given imaging task is estimated. For each application, a number of example tasks involving detection of an object of variable size and contrast against a noisy background are considered. The method described is fairly general and can be extended to account for a variety of imaging tasks. For the specific examples considered, the preliminary results estimate optimal CsI:Tl thicknesses of approximately 450 micrometer (approximately 200 mg/cm²), approximately 320 micrometer (approximately 140 mg/cm²), and approximately 200 micrometer (approximately 90 mg/cm²) for chest radiography, fluoroscopy, and mammography, respectively. These results are expected to depend upon the imaging task as well as upon the quality of available CsI:Tl, and future improvements in scintillator fabrication could result in increased optimal thickness and DQE.
No takes yet. Share an insight, caveat, or question.
Siewerdsen et al. (1998) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: