Key result
Ring-type cuffless BP device underestimates 24-hour systolic BP by ~7 mmHg vs conventional ABPM.
Why the study?
Cuffless blood pressure monitoring devices offer an alternative to conventional cuff-based ABPM, but their real-world performance across diurnal and nocturnal periods remains unclear.
Does a ring-type cuffless blood pressure device improve blood pressure measurement accuracy compared to conventional ambulatory blood pressure monitoring in adults?
Observational (n=154)
Does a ring-type cuffless blood pressure device improve blood pressure measurement accuracy compared to conventional ambulatory blood pressure monitoring in adults?
A ring-type cuffless blood pressure device systematically underestimated systolic BP and overestimated diastolic BP compared to conventional ABPM, limiting its current reliability for hypertension phenotyping.
Cuffless ring devices should not yet replace ABPM for phenotyping; leaves open validation needs in broader cohorts.
Objective: Cuffless blood pressure (BP) monitoring devices offer a potential alternative to conventional cuff-based ambulatory BP monitoring (ABPM), but their real-world performance across diurnal and nocturnal periods remains unclear; we therefore aimed to evaluate the agreement of a ring-type cuffless device (CART) with ABPM over 24-hour, daytime, and nighttime periods and its accuracy for nocturnal hypertension and dipping phenotypes.Design and method: We conducted a prospective simultaneous monitoring study in 154 of 168 enrolled adults (mean age 62.1 ± 14.2 years, 41.6% men) after excluding those with predefined clinical conditions or insufficient valid readings, in whom both conventional ABPM and a cuffless wearable device (CART) were worn over 24 hours. CART measurements were initiated after calibration to ABPM on the opposite arm. BP values were compared for 24-hour, daytime, and nighttime periods using mean differences and Bland–Altman analysis. Agreement for nocturnal hypertension and dipping status was evaluated using Cohen's kappa. Factors associated with close agreement (|24-h SBP difference| <5 mmHg) were analyzed using multivariable logistic regression. Results: CART systematically underestimated systolic BP and overestimated diastolic BP compared with ABPM. For 24-hour SBP, CART values were 7.1 mmHg lower than ABPM (SD of difference 10.6 mmHg), with only 33.1% of subjects meeting the <=5 mmHg accuracy criterion. Daytime SBP showed a similar bias (–7.8 mmHg), whereas nighttime SBP differences were smaller (–1.7 mmHg). In contrast, diastolic BP was consistently higher on CART, particularly at night (+4.8 mmHg). Classification agreement for nocturnal hypertension was moderate (k=0.516, McNemar p=0.324), while dipping pattern concordance was limited, with CART showing significantly attenuated dipping (mean 3.7% vs 7.8% by ABPM). In multivariable analysis, older age (OR 1.04 per year), male sex (OR 2.57), and lower BMI (OR 0.88 per kg/m2) independently predicted close SBP agreement, whereas higher heart rate was also associated with better concordance (OR 2.19). Conclusions: CART showed moderate agreement with ABPM, particularly for heart rate and nocturnal hypertension, but systematic systolic BP underestimation and attenuated dipping—especially in younger and higher-BMI individuals—limit its reliability for hypertension phenotyping, indicating that calibration and further validation are required before clinical substitution.
No takes yet. Share an insight, caveat, or question.
Shin et al. (2026) conducted an observational in Blood pressure monitoring (n=154). Ring-type cuffless blood pressure device (CART) vs. Ambulatory blood pressure monitoring (ABPM) was evaluated on Agreement of 24-hour systolic blood pressure (MD -7.1 mmHg). A ring-type cuffless blood pressure device systematically underestimated 24-hour systolic blood pressure by 7.1 mmHg compared with conventional ambulatory monitoring.