Key result
Conversion of BNP/NT-proBNP results between Biosite, Bayer, and Roche assays using regression relationships is not reliable, though concordance using manufacturers' cutoff values was acceptable.
Why the study?
Can BNP and NT-proBNP results be reliably converted between Biosite, Bayer, and Roche assays in emergency center patients?
Cross-Sectional (n=150)
Can BNP and NT-proBNP results be reliably converted between Biosite, Bayer, and Roche assays in emergency center patients?
BNP and NT-proBNP values obtained from different commercial assays cannot be reliably converted using mathematical formulas, though clinical concordance using manufacturer cutoffs is acceptable.
Avoid regression-based BNP/NT-proBNP conversions between assays; cross-sectional data leaves reliable harmonization open for prospective study.
This study determined whether, for patient monitoring, it is feasible to convert B-type natriuretic peptide (BNP) results obtained using Triage (BNP, Biosite, San Diego, CA), Centaur (BNP, Bayer Diagnostics, Tarrytown, NY), and Elecsys 2010 (N-terminal proBNP; Roche, Indianapolis, IN) assays. Concordance between assays and effects of renal impairment also were assessed. Samples were primarily from emergency center patients. Biosite testing was performed immediately; Bayer and Roche testing was performed later on plasma stored frozen (–30°C). Logistic regression relationships were as follows: Bayer = 0.57 Biosite + 23.1, n = 121, R2 = 0.85; Roche = 6.09 Biosite – 220.4 + 1,131.6 (if female), n = 131, R2 = 0.57; and Roche = 15.34 Bayer + 2,400.8, n = 150, R2 = 0.23. An increased serum creatinine level (≥2 mg/dL [≥177 μmol/L]) influenced the Roche results. We conclude the following from this preliminary study: (1) Results from one method cannot be converted reliably to another using regression relationships. (2) When using manufacturers’cutoff values, concordance between assays was acceptable. (3) Renal impairment affected Roche results.
No takes yet. Share an insight, caveat, or question.
Sykes et al. (2005) conducted a cross-sectional in Patients requiring BNP/NT-proBNP monitoring (n=150). Biosite, Bayer, and Roche BNP/NT-proBNP assays was evaluated on Feasibility of converting results between assays and concordance. Conversion of BNP/NT-proBNP results between Biosite, Bayer, and Roche assays using regression relationships is not reliable, though concordance using manufacturers' cutoff values was acceptable.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: