Key result
Intra-arterial BP recordings in ICU patients demonstrate ~21 mmHg mean systolic variation over 10 minutes.
Why the study?
The study was conducted to evaluate the magnitude of systolic and diastolic blood pressure variation over a few minutes using intra-arterial recordings.
Observational (n=51)
Significant short-term variability in intra-arterial blood pressure (mean 21 mmHg systolic) suggests that relying on single-point values for diagnosing and treating hypertension may be inadequate.
Answers are grounded in this paper and cite their sources.
May caution against single BP readings for ICU hypertension decisions; hypothesis-generating for prospective variability-adjusted outcome studies.
Objectives: The study presents the magnitude of variation of systolic and diastolic pressures over a few minutes from intra-arterial pressure recordings in 51 hemodynamically stable patients in an intensive care unit. Materials and Methods: Conscious surgical ICU patients, with arterial pressure cannulae placed as the standard of care, were recruited after obtaining informed consent. A validated data acquisition system was used to record intra-arterial blood pressure for a period of 10 minutes following a Fast-Flush test. Only those recordings with acceptable dynamic characteristics as per Gardner’s criteria were included in the analysis. Results: Three categories of systolic and diastolic pressure variability namely “beat-to-beat” variability, “Respiratory” variability, and “Total magnitude of variation” during the duration of recording, were reported for each of the 51 patients. The mean (± SD) of the total magnitude of variation for systolic and diastolic pressures in the study sample were 21 ± 9 mmHg and 14 ± 5 mmHg respectively. Conclusion: Given the magnitude of systolic and diastolic pressure variations over a few minutes, the validity of reporting single- point values for these pressures and using single-point cut-offs for diagnosis and treatment of hypertension must be re-evaluated.
No takes yet. Share an insight, caveat, or question.
Mukadam et al. (2023) conducted an observational in Hemodynamically stable patients in an intensive care unit (n=51). Intra-arterial blood pressure recording was evaluated on Total magnitude of variation for systolic and diastolic pressures. Intra-arterial blood pressure recordings over 10 minutes in ICU patients demonstrated a mean total magnitude of variation of 21 ± 9 mmHg for systolic and 14 ± 5 mmHg for diastolic pressures.
Synapse has enriched 10 related papers. Consider them for comparative context: