Key result
The saline calibration method for trans-cardiac conductance accurately determined parallel conductance offset volume with a nonsignificant bias of -2.5 ml compared to the zero-volume method.
The saline calibration method accurately determines parallel conductance offset volume for the trans-cardiac conductance method in measuring left ventricular volume, though substantial interanimal variability requires individual calibration.
Answers are grounded in this paper and cite their sources.
Saline calibration validates in animals; leaves open clinical translation for LV volume measurement.
The trans-cardiac conductance (TCC) method provides on-line left ventricular (LV) volume signals by determining the electrical conductance of blood in the LV by means of central venous and epithoracic electrodes. Conductive structures outside the LV blood pool cause a "parallel conductance" offset term (Vp) that can be determined by bolus injections of hypertonic saline in the pulmonary artery (Vp(saline)), which cause a transient increase in blood conductivity. This study in anesthetized sheep evaluates the accuracy of the saline calibration method and the variabilities of Vp between animals, between hemodynamic conditions and during the cardiac cycle. The conventional intra-cardiac conductance catheter method was used to obtain independent estimates of Vp by the zero-volume method (Vp(zero volume)). Mean baseline Vp(saline) and Vp(zerovolume) were 104 +/- 6 ml and 106 +/- 6 ml, respectively. Bland-Altman analysis showed a small nonsignificant bias (-2.5 ml) and narrow limits of agreement (4.6 ml). Vp was not significantly different between hemodynamic conditions (baseline, dobutamine, volume load, propranolol), but had a substantial interanimal variability (IAV) (38%). Average variations during the cardiac cycle were < 10% of mean Vp. We conclude that the saline method can be applied to determine Vp for TCC. IAV is substantial, so that Vp must be determined in each animal, but within-animal variability is relatively small.
No takes yet. Share an insight, caveat, or question.
Staal et al. (2003) studied Anesthetized sheep. Saline calibration method for trans-cardiac conductance vs. Intra-cardiac conductance catheter method (zero-volume method) was evaluated on Parallel conductance offset volume (Vp) (bias -2.5 ml, p=nonsignificant). The saline calibration method for trans-cardiac conductance accurately determined parallel conductance offset volume with a nonsignificant bias of -2.5 ml compared to the zero-volume method.
Synapse has enriched 10 related papers. Consider them for comparative context: