Key result
Algorithmic iFR pullback analysis matches expert human interpretation with ~89% agreement for PCI appropriateness.
Why the study?
Interpretation of instantaneous wave-free ratio pressure-wire pull back data can be complex and subjective.
Does algorithmic interpretation of iFR pressure-wire pull back data provide noninferior assessment of appropriateness and strategy for PCI compared to expert human interpretation in patients with stable coronary artery disease?
Cross-Sectional (n=640)
Single-blind
Yes
Does algorithmic interpretation of iFR pressure-wire pull back data provide noninferior assessment of appropriateness and strategy for PCI compared to expert human interpretation in patients with stable coronary artery disease?
Algorithmic interpretation of iFR pull back traces offers a standardized, reproducible alternative that is noninferior to expert human assessment and automatically corrects for pressure-wire drift.
Answers are grounded in this paper and cite their sources.
Supports algorithmic iFR interpretation as reproducible alternative to experts; leaves open prospective outcome validation before clinical adoption.
OBJECTIVES: The aim of this study was to investigate whether algorithmic interpretation (AI) of instantaneous wave-free ratio (iFR) pressure-wire pull back data would be noninferior to expert human interpretation. BACKGROUND: Interpretation of iFR pressure-wire pull back data can be complex and is subjective. METHODS: Fifteen human experts interpreted 1,008 iFR pull back traces (691 unique, 317 duplicate). For each trace, experts determined the hemodynamic appropriateness for percutaneous coronary intervention (PCI) and, in such cases, the optimal physiological strategy for PCI. The heart team (HT) interpretation was determined by consensus of the individual expert opinions. The same 1,008 pull back traces were also interpreted algorithmically. The coprimary hypotheses of this study were that AI would be noninferior to the interpretation of the median expert human in determining: 1) the hemodynamic appropriateness for PCI; and 2) the physiological strategy for PCI. RESULTS: Regarding the hemodynamic appropriateness for PCI, the median expert human demonstrated 89.3% agreement with the HT in comparison with 89.4% for AI (p < 0.01 for noninferiority). Across the 372 cases judged as hemodynamically appropriate for PCI according to the HT, the median expert human demonstrated 88.8% agreement with the HT in comparison with 89.7% for AI (p < 0.0001 for noninferiority). On reproducibility testing, the HT opinion itself changed 1 in 10 times for both the appropriateness for PCI and the physiological PCI strategy. In contrast, AI showed no change. CONCLUSIONS: AI of iFR pressure-wire pull back data was noninferior to expert human interpretation in determining both the hemodynamic appropriateness for PCI and the optimal physiological strategy for PCI.
No takes yet. Share an insight, caveat, or question.
Cook et al. (2019) conducted a cross-sectional in Stable coronary artery disease (n=640). Algorithmic interpretation of instantaneous wave-free ratio (iFR) pressure-wire pull back data vs. Median expert human interpretation was evaluated on Agreement with Heart Team consensus on the hemodynamic appropriateness for PCI (p=<0.01). Algorithmic interpretation of iFR pressure-wire pull back data was noninferior to median expert human interpretation for determining hemodynamic appropriateness for PCI.
Synapse has enriched 10 related papers. Consider them for comparative context: