Key result
Vessel fractional flow reserve (vFFR) provided high discrimination for coronary stenosis (AUC 0.80; 95% CI 0.70-0.90), which was comparable to nonhyperemic pressure ratios (p=0.42).
Why the study?
Does vFFR provide comparable diagnostic performance to NHPR for detecting functionally significant coronary stenosis (FFR ≤0.80) in patients with intermediate coronary stenosis?
Observational (n=74)
Does vFFR provide comparable diagnostic performance to NHPR for detecting functionally significant coronary stenosis (FFR ≤0.80) in patients with intermediate coronary stenosis?
Effect estimate: AUC 0.80 (95% CI 0.70-0.90)
p-value: p=0.42
vFFR derived from 3D quantitative coronary angiography provides excellent diagnostic performance comparable to NHPR for detecting functionally significant coronary stenosis, potentially reducing the need for invasive pressure wire assessment.
May reduce pressure-wire use in intermediate lesions; leaves open prospective validation before practice change.
Background Fractional flow reserve (FFR) and nonhyperemic pressure ratios (NHPRs) have been widely used to assess the functional severity of coronary stenosis. However, their measurement requires using a pressure wire, making their use in all patients difficult. The recently developed vessel fractional flow reserve (vFFR), derived from three‐dimensional quantitative coronary angiography, is expected to serve as a surrogate for pressure wire assessment. Methods This retrospective study was conducted on patients with intermediate coronary stenosis who underwent FFR and NHPR measurements. The vFFR and NHPR values were compared for diagnosing coronary stenosis as defined by an FFR of ≤0.80, and the number of patients not requiring wire‐based assessment was estimated. Results In a total of 90 lesions from 74 patients (median [SD] age 75 [12] years; men 80%), the median FFR was 0.78 (0.72–0.84), and 57% of these lesions ( N = 51) exhibited an FFR of ≤0.80. vFFR provided high discrimination for coronary stenosis (area under the curve 0.80, 95% confidence interval 0.70–0.90), which was comparable to that of NHPRs ( p = 0.42). High diagnostic accuracy was consistently observed across a variety of clinical presentations (i.e., old age, diabetes, target coronary artery, and left ventricular hypertrophy) ( p interaction > 0.05). In total, 55 lesions (61%) demonstrated positive or negative likelihood of coronary stenosis when vFFR was <0.73 (specificity 90%) or >0.87 (sensitivity 88%), respectively. Conclusion vFFR demonstrated excellent diagnostic performance for detecting functionally significant coronary stenosis as evaluated by FFR. vFFR may be used as a surrogate for pressure wire assessment.
No takes yet. Share an insight, caveat, or question.
Ikeda et al. (2022) conducted an observational in Intermediate coronary stenosis (n=74). Vessel fractional flow reserve (vFFR) vs. Nonhyperemic pressure ratios (NHPRs) was evaluated on Diagnosing coronary stenosis as defined by an FFR of ≤0.80 (AUC 0.80, 95% CI 0.70-0.90, p=0.42). Vessel fractional flow reserve (vFFR) provided high discrimination for coronary stenosis (AUC 0.80; 95% CI 0.70-0.90), which was comparable to nonhyperemic pressure ratios (p=0.42).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: