Key result
MobiusHD EVBA device lowers office blood pressure by ~24/12 mmHg in resistant hypertension.
Why the study?
Disappointing sham-controlled renal nerve denervation trials created an unmet need to explore alternative medical device approaches for resistant hypertension.
Does the MobiusHD EVBA device improve blood pressure and maintain safety in adults with resistant hypertension?
Does the MobiusHD EVBA device improve blood pressure and maintain safety in adults with resistant hypertension?
Endovascular carotid baroreflex amplification with the MobiusHD device may safely lower blood pressure in resistant hypertension, laying the groundwork for ongoing randomized sham-controlled trials.
Hypothesis-generating for resistant hypertension; leaves open safety and efficacy confirmation in randomized sham-controlled trials.
Introduction Early promising results of renal nerve denervation awakened interest in developing medical device alternatives for patients with resistant hypertension. The subsequent sham-controlled renal nerve denervation randomized trials were disappointing leading researchers and innovators to explore alternative device and trial designs to address this significant unmet need. We describe the innovation process leading to the first endovascular carotid baroreflex amplification device currently undergoing clinical trials in the United States and Europe.Areas covered We provide a brief overview of carotid baroreceptor physiology and then couple this knowledge with the fundamental principles of strain pattern changes that led to the proposed innovation. The mechanism of blood pressure reduction via enhancing innate physiologic carotid sinus baroreceptor signaling through changes in pulsatile focal carotid bulb strain is described alongside preclinical testing and early clinical results.Expert opinion The collective data to date suggest endovascular carotid baroreflex amplification may be an innovative alternative for resistant hypertension patients. However, well-controlled studies will be needed to assess efficacy, safety, durability, and risk: benefit of this permanent intravascular carotid implant.
No takes yet. Share an insight, caveat, or question.
Bates et al. (2020) conducted a review in resistant hypertension (n=30). MobiusHD Endovascular Baroreflex Amplification (EVBA) device was evaluated on Incidence of serious adverse events at 6 months.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: