Key result
CMR answers clinical questions in only ~20% of pediatric patients with S-ICDs due to LV artifact.
Why the study?
Does the presence of a subcutaneous implantable cardioverter-defibrillator reduce CMR image quality and diagnostic utility in pediatric patients?
Case Report (n=4)
No
Does the presence of a subcutaneous implantable cardioverter-defibrillator reduce CMR image quality and diagnostic utility in pediatric patients?
In pediatric patients with SICDs, CMR image quality is significantly degraded by artifact over the left ventricle, though right ventricular evaluation may still remain feasible.
Answers are grounded in this paper and cite their sources.
May limit CMR utility in pediatric SICD patients; leaves open need for alternative imaging strategies.
Key Teaching Points•Subcutaneous implantable cardioverter-defibrillators (SICDs) can hamper the utility of cardiac magnetic resonance imaging (CMR) owing to image artifact they can produce. Depending on the clinical question, patient and diagnosis-specific imaging protocols may be feasible.•For those patients requiring evaluation of cardiac function and myocardial scarring, it may be prudent to undergo CMR prior to implantation of an SICD, or more generally, prior to implantation of any cardiovascular implantable electronic devices.•In pediatric patients with SICDs, significant distortion and artifact were notable over the left ventricle, preventing the CMR from answering the clinical questions for which it was ordered. The right ventricle was relatively spared in terms of visualization—in this scenario, CMR may have utility if the clinical questions are more directed toward right ventricular function or right ventricular myocardial scarring.•The imaging examples provided offer a glimpse into the type and extent of distortion an SICD can impose upon CMR images and may change the pediatric cardiologists’ clinical practice in terms of timing or indication for ordering CMR in these patients.•There were no adverse outcomes or device malfunctions immediately post CMR in the 5 cases with SICDs. •Subcutaneous implantable cardioverter-defibrillators (SICDs) can hamper the utility of cardiac magnetic resonance imaging (CMR) owing to image artifact they can produce. Depending on the clinical question, patient and diagnosis-specific imaging protocols may be feasible.•For those patients requiring evaluation of cardiac function and myocardial scarring, it may be prudent to undergo CMR prior to implantation of an SICD, or more generally, prior to implantation of any cardiovascular implantable electronic devices.•In pediatric patients with SICDs, significant distortion and artifact were notable over the left ventricle, preventing the CMR from answering the clinical questions for which it was ordered. The right ventricle was relatively spared in terms of visualization—in this scenario, CMR may have utility if the clinical questions are more directed toward right ventricular function or right ventricular myocardial scarring.•The imaging examples provided offer a glimpse into the type and extent of distortion an SICD can impose upon CMR images and may change the pediatric cardiologists’ clinical practice in terms of timing or indication for ordering CMR in these patients.•There were no adverse outcomes or device malfunctions immediately post CMR in the 5 cases with SICDs.
No takes yet. Share an insight, caveat, or question.
Gaffar et al. (2022) conducted a case report in Cardiomyopathy with subcutaneous implantable cardioverter-defibrillators (SICDs) (n=4). Subcutaneous implantable cardioverter-defibrillator (SICD) was evaluated on Ability of cardiac magnetic resonance imaging to answer the clinical question. Cardiac magnetic resonance imaging was able to answer the clinical question in only 20% of studies in pediatric patients with subcutaneous implantable cardioverter-defibrillators due to severe artifact over the left ventricle.
Synapse has enriched 10 related papers. Consider them for comparative context: