Why the study?
Does the bin area method (BAM) compare favorably to correlation waveform analysis (CWA) for discriminating retrograde from anterograde atrial activation?
Does the bin area method (BAM) compare favorably to correlation waveform analysis (CWA) for discriminating retrograde from anterograde atrial activation?
The bin area method is as effective as correlation waveform analysis for discriminating retrograde from anterograde atrial activation but is significantly more computationally efficient.
Answers are grounded in this paper and cite their sources.
BAM may enable efficient atrial activation analysis; leaves open clinical adoption pending larger validation studies.
Accurate detection of pacemaker-mediated tachycardia requires a rapid and reliable method for discriminating retrograde atrial activation (RAA) from anterograde atrial activation (AAA). The authors compared two methods of morphology analysis for discriminating RAA from AAA applied to the atrial electrogram: correlation waveform analysis (CWA) and the bin area method (BAM). Both methods were tested on 19 patients undergoing clinical electrophysiology studies. Both CWA and BAM discriminated AAA from 1:1 RAA in 19/19 patients, but BAM reduced the computational demands by a factor of 6.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
No takes yet. Share an insight, caveat, or question.
Throne et al. (2003) studied this question.