Why the study?
What are the radiographic, CT, and echocardiographic findings of complications associated with automatic implantable cardioverter defibrillators?
What are the radiographic, CT, and echocardiographic findings of complications associated with automatic implantable cardioverter defibrillators?
Distorted AICD patches may indicate postoperative complications, and CT is the preferred imaging modality for their evaluation.
Answers are grounded in this paper and cite their sources.
Distorted AICD patches may flag complications on CT; leaves open optimal surveillance strategies.
Automatic implantable cardioverter defibrillators (AICDs) were studied in three groups: (a) Serial radiographs were reviewed in 51 clinic patients. Twenty of 96 (21%) AICD patches distorted with time. (b) Thirty-six postoperative computed tomographic (CT) scans of asymptomatic patients revealed that pericardial fluid collections were frequent during the month after surgery but rare beyond that. Echocardiography was insensitive for these collections. CT also demonstrated dense fibrosis around some distorted patches, months after surgery. (c) Five other patients with pericardial infection had distorted patches, and the four studied with CT had fluid beneath their patches. (d) A case of constrictive pericarditis had distorted patches but was not diagnosed with CT. The authors conclude that distorted patches may indicate postoperative complications and that CT is the imaging modality of choice.