Key result
Constrictive pericarditis occurred in 8.6% (3 of 35) of patients who received ICDs with patch electrodes during a mean follow-up of 24 months, resolving after patch removal.
Why the study?
Does ICD implantation with patch electrodes cause constrictive pericarditis in patients receiving ICDs?
Observational (n=35)
Does ICD implantation with patch electrodes cause constrictive pericarditis in patients receiving ICDs?
Answers are grounded in this paper and cite their sources.
Constrictive pericarditis is a potential complication of epicardial patch electrodes for ICDs and should be considered in patients presenting with cardiac decompensation.
Constrictive pericarditis can be associated with ICD patch electrodes. During a mean follow-up of 24 months, in a population of 35 patients who received ICDs with a patch electrodes configuration, we identified three patients with clinical and hemodynamic signs compatible with this event. Patient 1, a 35-year-old male, underwent implantation of an ICD because of a primary electrical disease complicated by cardiac arrest. Fourteen months later he complained of exertional dyspnea without any signs of heart failure. Right heart catheterization showed high filling pressures and diastolic dip and plateau in pressure curves. Thoracotomy and pericardial exploration were performed. Three months after removal of the patches and insertion of an endocardial lead system, the patient had normal respiration. Patients 2 and 3, who suffered from coronary heart disease without heart failure, exhibited a hemodynamic profile suggestive of constrictive pericarditis: in one patient, 10 months after ICD implantation, associated with right heart failure; and in the other, 18 months after ICD implantation with left heart failure. Patch electrodes were removed in these two patients and replaced by endocardial lead electrodes with subsequent clinical improvement. It is concluded that constrictive pericarditis related to epicardial patch is not an uncommon occurrence during ICD therapy and should be considered in patients who show clinical signs of cardiac decompensation.
No takes yet. Share an insight, caveat, or question.
Chevalier et al. (1996) conducted an observational in Patients receiving ICDs with patch electrodes (n=35). ICD patch electrodes was evaluated on Clinical and hemodynamic signs compatible with constrictive pericarditis. Constrictive pericarditis occurred in 8.6% (3 of 35) of patients who received ICDs with patch electrodes during a mean follow-up of 24 months, resolving after patch removal.
Synapse has enriched 10 related papers. Consider them for comparative context: