Key result
Late right ventricular perforation after transvenous active-fixation ICD lead implantation presented as recurrent pericardial chest pain and was confirmed by right ventriculography.
Case Report (n=1)
Late right ventricular perforation should be considered in patients with recurrent pericardial chest pain after active-fixation ICD lead implantation, even if initial pacing parameters are normal.
Answers are grounded in this paper and cite their sources.
Consider late RV perforation in recurrent pericardial pain after active-fixation ICD leads; case report leaves open incidence and management.
KAUTZNER, J., et al.: Recurrent Pericardial Chest Pain: A Case of Late Right Ventricular Perforation After Implantation of a Transvenous Active‐Fixation ICD Lead. A 36‐year‐old woman with a history of recurrent syncopal episodes presumably due to ventricular tachyarrhythmia in mitral valve prolapse underwent implantation of a transvenous ICD system. During a 23‐month follow‐up, she developed recurrent pericardial chest pain with pericardial friction rub. The first episode of chest pain occurred without any detectable change in pacing or sensing parameters. The second episode was associated with an increase in pacing threshold and drop in intracardiac signal amplitude. Right ventricular perforation was suspected fluoroscopically and confirmed by right ventriculography. This case report emphasizes the key steps in the diagnosis of this rare complication of an ICD implantation.
No takes yet. Share an insight, caveat, or question.
Kautzner et al. (2001) conducted a case report in Recurrent pericardial chest pain after ICD implantation (n=1). Transvenous active-fixation ICD lead implantation was evaluated on Right ventricular perforation. Late right ventricular perforation after transvenous active-fixation ICD lead implantation presented as recurrent pericardial chest pain and was confirmed by right ventriculography.
Synapse has enriched 10 related papers. Consider them for comparative context: