Key result
The k-t GRAPPA accelerated 4D-flow sequence reduced mean scan time by 46% compared to standard GRAPPA while providing fewer motion artefacts and comparable flow quantification.
Why the study?
The study aimed to compare two broadly used accelerated 4D-flow sequences with a 2D-flow reference standard in healthy volunteers regarding absolute flow parameters, image quality, and eddy current correction.
Does a k-t GRAPPA accelerated 4D-flow MRI sequence improve scan time and image quality compared to standard GRAPPA without compromising aortic flow quantification accuracy?
Cross-Sectional (n=40)
Blinded reading
Randomized order of reading
No
Does a k-t GRAPPA accelerated 4D-flow MRI sequence improve scan time and image quality compared to standard GRAPPA without compromising aortic flow quantification accuracy?
k-t GRAPPA accelerated 4D-flow MRI significantly reduces scan time and motion artifacts compared to standard GRAPPA while maintaining high accuracy for aortic flow quantification.
Answers are grounded in this paper and cite their sources.
May improve 4D-flow MRI feasibility in busy clinics; leaves open multicenter validation of diagnostic impact.
To compare two broadly used 4D-flow- with a 2D-flow-sequence in healthy volunteers, regarding absolute flow parameters, image quality (IQ), and eddy current correction (ECC). Forty volunteers (42 ± 11.8 years, 22 females) were examined with a 3T scanner. Thoracic aortic flow was assessed using a 3D-T2w-SPACE-STIR-sequence for morphology and two accelerated 4D-flow sequences for comparison, one with k-t undersampling and one with standard GRAPPA parallel-imaging. 2D-flow was used as reference standard. The custom-made software tool Bloodline enabled flow measurements for all analyses at the same location. Quantitative flow analyses were performed with and without ECC. One reader assessed pathline IQ (IQ-PATH) and occurrence of motion artefacts (IQ-ART) on a 3-point grading scale, the higher the better. k-t GRAPPA allowed a significant mean scan time reduction of 46% (17:56 ± 5:26 min vs. 10:40 ± 3:15 min) and provided significantly fewer motion artefacts than standard GRAPPA (IQ-ART 1.57 ± 0.55 vs. 0.84 ± 0.48; p < 0.001). Neither 4D-flow sequence significantly differed in flow volume nor peak velocity results with or without ECC. Nevertheless, the correlation between both 4D-flow sequences and 2D-flow was better with ECC; the k-t GRAPPA sequence performed best (R = 0.96 vs. 0.90). k-t GRAPPA 4D-flow was not inferior to a standard GRAPPA-sequence, showed fewer artefacts, comparable IQ and was almost two-fold faster.
No takes yet. Share an insight, caveat, or question.
Ebel et al. (2019) conducted a cross-sectional in Healthy volunteers (n=40). k-t GRAPPA accelerated 4D-flow sequence vs. Standard GRAPPA 4D-flow sequence was evaluated on Mean acquisition time (46.12% reduction, p=0.002). The k-t GRAPPA accelerated 4D-flow sequence reduced mean scan time by 46% compared to standard GRAPPA while providing fewer motion artefacts and comparable flow quantification.