Key result
Disabling ICD respiratory rate trend diagnostics resolves device-lead mismatch atrial noise.
Why the study?
Does the respiratory rate trend diagnostic cause atrial lead noise in ICD patients?
Case Report (n=1)
Does the respiratory rate trend diagnostic cause atrial lead noise in ICD patients?
The respiratory rate trend diagnostic in Boston Scientific ICDs can cause intermittent high-frequency noise and should be disabled if unexplained oversensing or impedance issues occur.
Answers are grounded in this paper and cite their sources.
In ICD patients with lead mismatch and unexplained atrial noise, consider disabling respiratory rate trend; single-case data leaves generalizability open.
Key Teaching Points•In Boston Scientific devices, the respiratory rate trend heart failure diagnostic should be disabled in those patients who have intermittent high-frequency noise seen on either lead, as well as out-of-range impedances that are otherwise unexplained.•The likelihood of minute ventilation sensor signal oversensing is higher in patients in whom there is a mismatch between the device can and the lead.•This artifact can be seen not only in pacemakers but in implantable cardioverter-defibrillators as well. •In Boston Scientific devices, the respiratory rate trend heart failure diagnostic should be disabled in those patients who have intermittent high-frequency noise seen on either lead, as well as out-of-range impedances that are otherwise unexplained.•The likelihood of minute ventilation sensor signal oversensing is higher in patients in whom there is a mismatch between the device can and the lead.•This artifact can be seen not only in pacemakers but in implantable cardioverter-defibrillators as well.
No takes yet. Share an insight, caveat, or question.
McClelland et al. (2018) conducted a case report in Implantable cardioverter-defibrillator with intermittent atrial lead noise (n=1). Minute ventilation sensor for respiratory rate trending vs. Disabling the sensor was evaluated on Resolution of atrial lead noise and high impedance. In a patient with an implantable cardioverter-defibrillator and device-lead mismatch, the respiratory rate trend heart failure diagnostic caused intermittent high-frequency atrial lead noise that resolved upon disabling the feature.
Synapse has enriched 10 related papers. Consider them for comparative context: